Izewska Joanna, Georg Dietmar, Bera Pranabes, Thwaites David, Arib Mehenna, Saravi Margarita, Sergieva Katia, Li Kaibao, Yip Fernando Garcia, Mahant Ashok Kumar, Bulski Wojciech
Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, Vienna, Austria.
Radiother Oncol. 2007 Jul;84(1):67-74. doi: 10.1016/j.radonc.2007.06.006.
A strategy for national TLD audit programmes has been developed by the International Atomic Energy Agency (IAEA). It involves progression through three sequential dosimetry audit steps. The first step audits are for the beam output in reference conditions for high-energy photon beams. The second step audits are for the dose in reference and non-reference conditions on the beam axis for photon and electron beams. The third step audits involve measurements of the dose in reference, and non-reference conditions off-axis for open and wedged symmetric and asymmetric fields for photon beams. Through a co-ordinated research project the IAEA developed the methodology to extend the scope of national TLD auditing activities to more complex audit measurements for regular fields.
Based on the IAEA standard TLD holder for high-energy photon beams, a TLD holder was developed with horizontal arm to enable measurements 5cm off the central axis. Basic correction factors were determined for the holder in the energy range between Co-60 and 25MV photon beams.
New procedures were developed for the TLD irradiation in hospitals. The off-axis measurement methodology for photon beams was tested in a multi-national pilot study. The statistical distribution of dosimetric parameters (off-axis ratios for open and wedge beam profiles, output factors, wedge transmission factors) checked in 146 measurements was 0.999+/-0.012.
The methodology of TLD audits in non-reference conditions with a modified IAEA TLD holder has been shown to be feasible.
国际原子能机构(IAEA)制定了一项国家热释光剂量计(TLD)审核计划策略。该策略包括三个连续的剂量测定审核步骤。第一步审核针对高能光子束在参考条件下的射束输出。第二步审核针对光子束和电子束在射束轴上参考和非参考条件下的剂量。第三步审核涉及对光子束开放和楔形对称及不对称射野在参考和非参考条件下离轴剂量的测量。通过一个协调研究项目,IAEA开发了将国家TLD审核活动范围扩展至常规射野更复杂审核测量的方法。
基于IAEA高能光子束标准TLD支架,开发了一种带有水平臂的TLD支架,以实现离中心轴5厘米处的测量。确定了该支架在钴 - 60至25兆伏光子束能量范围内的基本校正因子。
为医院中的TLD辐照制定了新程序。光子束离轴测量方法在一项多国试点研究中进行了测试。在146次测量中检查的剂量学参数(开放和楔形射野轮廓的离轴比、输出因子、楔形透射因子)的统计分布为0.999±0.012。
已证明使用改进的IAEA TLD支架在非参考条件下进行TLD审核的方法是可行的。