Followill David S, Evans DeeAnn Radford, Cherry Christopher, Molineu Andrea, Fisher Gary, Hanson William F, Ibbott Geoffrey S
Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030, USA.
Med Phys. 2007 Jun;34(6):2070-6. doi: 10.1118/1.2737158.
The Radiological Physics Center (RPC) developed two heterogeneous anthropomorphic quality assurance phantoms for use in verifying the accuracy of radiation delivery: one for intensity-modulated radiation therapy (IMRT) to the pelvis and the other for stereotactic body radiation therapy (SBRT) to the thorax. The purpose of this study was to describe the design and development of these two phantoms and to demonstrate the reproducibility of measurements generated with them. The phantoms were built to simulate actual patient anatomy. They are lightweight and water-fillable, and they contain imageable targets and organs at risk of radiation exposure that are of similar densities to their human counterparts. Dosimetry inserts accommodate radiochromic film for relative dosimetry and thermoluminesent dosimetry capsules for absolute dosimetry. As a part of the commissioning process, each phantom was imaged, treatment plans were developed, and radiation was delivered at least three times. Under these controlled irradiation conditions, the reproducibility of dose delivery to the target TLD in the pelvis and thorax phantoms was 3% and 0.5%, respectively. The reproducibility of radiation-field localization was less than 2.5 mm for both phantoms. Using these anthropomorphic phantoms, pelvic IMRT and thoracic SBRT radiation treatments can be verified with a high level of precision. These phantoms can be used to effectively credential institutions for participation in specific NCI-sponsored clinical trials.
放射物理中心(RPC)开发了两种用于验证放射治疗准确性的非均匀人体模体:一种用于盆腔调强放射治疗(IMRT),另一种用于胸部立体定向体部放射治疗(SBRT)。本研究的目的是描述这两种体模的设计与开发,并展示使用它们所产生测量结果的可重复性。这些体模被制作成模拟实际患者的解剖结构。它们重量轻且可注水,包含可成像的靶区和可能受到辐射暴露的危险器官,其密度与人体相应部位相似。剂量学插入物可容纳用于相对剂量测定的放射变色胶片和用于绝对剂量测定的热释光剂量测定胶囊。作为调试过程的一部分,对每个体模进行成像,制定治疗计划,并至少进行三次放射治疗。在这些受控的照射条件下,盆腔和胸部体模中靶区热释光剂量计的剂量传递可重复性分别为3%和0.5%。两种体模的辐射野定位可重复性均小于2.5毫米。使用这些人体模体,可以高精度地验证盆腔IMRT和胸部SBRT放射治疗。这些体模可用于有效地认证机构参与特定的由美国国立癌症研究所赞助的临床试验。