Blaschko Sarah D, Deane Leslie Allan, Krebs Alfred, Abdelshehid Corollos S, Khan Farhan, Borin James, Nguyen Alex, McDougall Elspeth M, Clayman Ralph V
Department of Urology, University of California-Irvine Medical Center, Orange 92868, California, USA.
J Endourol. 2007 Jul;21(7):780-3. doi: 10.1089/end.2006.0315.
To characterize the flow of a novel ureteral stent composed of a nickel-cobalt-chromium-molybde-num alloy and compare it with flow in a standard ureteral stent.
Six 6F Resonance stents and six 6F standard Black Beauty ureteral stents were placed in six Yucatan minipigs, with each pig serving as its own control. Flow assessment was performed on all stents via a nephrostomy tube delivering a standard rate of 0.9% saline at 35 cm H(2)O. Flow studies on the standard stents encompassed extraluminal (i.e., lumen of stent occluded with a guidewire), intraluminal (i.e., ureter secured to stent with a constricting suture), and combined (i.e., open lumen without constricting suture) flow. In the Resonance stent, only combined and intraluminal flow could be addressed, as there is no access to the lumen of this stent.
With the Resonance stent, intraluminal flow was much greater than combined flow, with mean values of 5.15 mL/min and 2.50 mL/min, respectively (P = 0.057; SD = 7.73). Intraluminal flow was similar to combined flow in the 6F standard stent, with mean values of 7.34 mL/min and 7.30 mL/min, respectively (P = 0.88; SD = 1.76). The standard stent had significantly greater combined flow than the Resonance stent (P = 0.023) but not intraluminal flow (P = 0.247). Of note, whereas it was possible to occlude the 6F standard stent completely with a ureteral ligature (i.e., no guidewire placed in the lumen), it was not possible to occlude the Resonance stent regardless of how tightly the suture was tied.
The Resonance metal alloy stent provides less overall flow than a standard stent. However, under circumstances of extrinsic ureteral compression sufficient to occlude a standard stent (e.g., extrinsic compression plus an internal guidewire), the metal stent continues to provide satisfactory drainage.