University of Texas Health Science Center at San Antonio, San Antonio, Texas 78229, USA.
Med Phys. 2007 Jul;34(7):2949-56. doi: 10.1118/1.2748105.
Quality assurance (QA) of the multileaf collimator (MLC) is a critical step for the delivery of intensity modulated radiation therapy treatment plan. While QA procedures for motor-driven MLC have been published extensively, those for binary MLCs such as the one used for helical tomotherapy have not been presented in the literature, as this is still a fairly new technology. In this study, seven test patterns for the MLC QA of a helical tomotherapy unit have been designed and implemented. The seven test patterns check the MLC alignment, MLC leakage, MLC timing and MLC leaf position error in detail. Those patterns can be easily implemented in any center with a helical tomotherapy unit as part of the routine QA. The QA procedures can be performed using existing QA resources such as solid water phantom and EDR2 film. A software toolkit called "Tomo MLC QA" has been developed to assist in generating the QA procedures and analyzing the results. Our results showed that the helical tomotherapy MLC is very robust, exhibiting interleaf leakage of 0.53% +/-0.09%. Several issues with the MLC have been found and discussed. The QA results also illustrate the utilization and usefulness of the proposed QA procedures.
质量保证(QA)多叶准直器(MLC)是调强放射治疗计划交付的关键步骤。虽然已经广泛发布了用于电动 MLC 的 QA 程序,但对于诸如螺旋断层放疗中使用的二进制 MLC 等技术尚未在文献中介绍,因为这仍然是一项相当新的技术。在这项研究中,已经设计并实施了七种用于螺旋断层放疗设备的 MLC QA 测试模式。这七种测试模式详细检查了 MLC 的对准、MLC 的泄漏、MLC 的定时和 MLC 叶片位置误差。这些模式可以作为常规 QA 的一部分,很容易在任何配备螺旋断层放疗设备的中心实施。QA 程序可以使用现有的 QA 资源(如水凝胶体模和 EDR2 胶片)进行。已经开发了一个名为“Tomo MLC QA”的软件工具包,以协助生成 QA 程序和分析结果。我们的结果表明,螺旋断层放疗 MLC 非常稳健,表现出 0.53% +/-0.09%的叶间泄漏。已经发现并讨论了几个 MLC 问题。QA 结果还说明了所提出的 QA 程序的利用和有用性。