使用近距离放射治疗机器人进行旋转针插入的观察

Observations on rotating needle insertions using a brachytherapy robot.

作者信息

Meltsner M A, Ferrier N J, Thomadsen B R

机构信息

Department of Medical Physics, University of Wisconsin, Madison, WI 53706, USA.

出版信息

Phys Med Biol. 2007 Oct 7;52(19):6027-37. doi: 10.1088/0031-9155/52/19/021. Epub 2007 Sep 17.


DOI:10.1088/0031-9155/52/19/021
PMID:17881817
Abstract

A robot designed for prostate brachytherapy implantations has the potential to greatly improve treatment success. Much of the research in robotic surgery focuses on measuring accuracy. However, there exist many factors that must be optimized before an analysis of needle placement accuracy can be determined. Some of these parameters include choice of the needle type, insertion velocity, usefulness of the rotating needle and rotation speed. These parameters may affect the force at which the needle interacts with the tissue. A reduction in force has been shown to decrease the compression of the prostate and potentially increase the accuracy of seed position. Rotating the needle as it is inserted may reduce frictional forces while increasing accuracy. However, needle rotations are considered to increase tissue damage due to the drilling nature of the insertion. We explore many of the factors involved in optimizing a brachytherapy robot, and the potential effects each parameter may have on the procedure. We also investigate the interaction of rotating needles in gel and suggest the rotate-cannula-only method of conical needle insertion to minimize any tissue damage while still maintaining the benefits of reduced force and increased accuracy.

摘要

一款专为前列腺近距离放射治疗植入设计的机器人有潜力极大地提高治疗成功率。机器人手术的许多研究都集中在测量准确性上。然而,在确定针放置准确性的分析之前,存在许多必须优化的因素。其中一些参数包括针类型的选择、插入速度、旋转针的效用和转速。这些参数可能会影响针与组织相互作用的力。已表明力的降低会减少前列腺的压缩,并可能提高种子位置的准确性。在插入针时旋转针可能会减少摩擦力,同时提高准确性。然而,由于插入的钻孔性质,针的旋转被认为会增加组织损伤。我们探讨了优化近距离放射治疗机器人所涉及的许多因素,以及每个参数可能对手术产生的潜在影响。我们还研究了旋转针在凝胶中的相互作用,并建议采用仅旋转套管的锥形针插入方法,以尽量减少任何组织损伤,同时仍保持降低力和提高准确性的益处。

相似文献

[1]
Observations on rotating needle insertions using a brachytherapy robot.

Phys Med Biol. 2007-10-7

[2]
Development of a tapping device: a new needle insertion method for prostate brachytherapy.

Phys Med Biol. 2006-2-21

[3]
Robot-assisted minimally invasive lung brachytherapy.

Int J Med Robot. 2007-3

[4]
Teleoperated master-slave needle insertion.

Int J Med Robot. 2009-12

[5]
Robot-assisted prostate brachytherapy.

Med Image Comput Comput Assist Interv. 2006

[6]
Measurement of prostate rotation during insertion of needles for brachytherapy.

Radiother Oncol. 2005-12

[7]
A Hand-Held Assistant for Semiautomated Percutaneous Needle Steering.

IEEE Trans Biomed Eng. 2017-3

[8]
Design of an ultrasound-guided robotic brachytherapy needle-insertion system.

Annu Int Conf IEEE Eng Med Biol Soc. 2009

[9]
Control of soft tissue deformation during robotic needle insertion.

Minim Invasive Ther Allied Technol. 2006

[10]
Needle insertion into soft tissue: a survey.

Med Eng Phys. 2007-5

引用本文的文献

[1]
Experimental Study of Needle Insertion into Gerbil Tympanic Membrane.

J Assoc Res Otolaryngol. 2024-10

[2]
Design and evaluation of a pneumatic actuation unit for a wasp-inspired self-propelled needle.

PLoS One. 2024

[3]
In-Silico Analysis of Optimal Configurations for Rotational Bioinspired Bone Marrow Biopsy Needle Designs: An ANN Approach.

Ann Biomed Eng. 2024-4

[4]
Design and Experiment of Assistive Mechanism for Adjustment of Transrectal Ultrasound Probe.

Appl Bionics Biomech. 2020-10-15

[5]
Design of an ultra-thin steerable probe for percutaneous interventions and preliminary evaluation in a gelatine phantom.

PLoS One. 2019-9-4

[6]
Force Modeling, Identification, and Feedback Control of Robot-Assisted Needle Insertion: A Survey of the Literature.

Sensors (Basel). 2018-2-12

[7]
A novel approach to an automated needle insertion in brachytherapy procedures.

Med Biol Eng Comput. 2017-7-15

[8]
Hyper- and viscoelastic modeling of needle and brain tissue interaction.

Annu Int Conf IEEE Eng Med Biol Soc. 2014

[9]
Robotic ultrasound and needle guidance for prostate cancer management: review of the contemporary literature.

Curr Opin Urol. 2014-1

[10]
Evaluation of a robotic technique for transrectal MRI-guided prostate biopsies.

Eur Radiol. 2011-9-29

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

推荐工具

医学文档翻译智能文献检索