Magn Reson Imaging. 2007 Oct;25(8):1138-47. doi: 10.1016/j.mri.2007.01.004. Epub 2007 Feb 20.

Abstract

Coronary magnetic resonance angiography (MRA) acquired using steady-state free precession (SSFP) sequences tends to suffer from image artifacts caused by local magnetic field inhomogeneities. Flow- and gradient-switching-induced eddy currents are important sources of such phase errors, especially under off-resonant conditions. In this study, we propose to reduce these image artifacts by using a linear centric-encoding (LCE) scheme in the phase-encoding (PE) direction. Abrupt change in gradients, including magnitude and polarity between consecutive radiofrequency cycles, is minimized using the LCE scheme. Results from numeric simulations and phantom studies demonstrated that signal oscillation can be markedly reduced using LCE as compared to conventional alternating centric-encoding (ACE) scheme. The image quality of coronary arteries was improved at both 1.5 and 3.0 T using LCE compared to those acquired using ACE PE scheme (1.5 T: ACE/LCE=2.2+/-0.8/3.0+/-0.6, P=.02; 3.0 T: ACE/LCE=2.1+/-1.1/3.0+/-0.8, P=.01). In conclusion, flow- and eddy-currents-induced imaging artifacts in coronary MRA using SSFP sequence can be markedly reduced with LCE acquisition of PE lines.

摘要

使用稳态自由进动(SSFP)序列采集的冠状动脉磁共振血管造影(MRA)往往会受到局部磁场不均匀性引起的图像伪影影响。流动和梯度切换引起的涡流是此类相位误差的重要来源,尤其是在非共振条件下。在本研究中,我们建议通过在相位编码(PE)方向上使用线性中心编码(LCE)方案来减少这些图像伪影。使用LCE方案可将连续射频周期之间包括幅度和极性在内的梯度突变降至最低。数值模拟和体模研究结果表明,与传统的交替中心编码(ACE)方案相比,使用LCE可显著减少信号振荡。与使用ACE PE方案采集的图像相比,在1.5 T和3.0 T场强下使用LCE均改善了冠状动脉的图像质量(1.5 T:ACE/LCE = 2.2±0.8/3.0±0.6,P = 0.02;3.0 T:ACE/LCE = 2.1±1.1/3.0±0.8,P = 0.01)。总之,采用LCE采集PE线可显著减少使用SSFP序列的冠状动脉MRA中流动和涡流引起的成像伪影。

相似文献

3
Analysis and compensation of eddy currents in balanced SSFP.
Magn Reson Med. 2005 Jul;54(1):129-37. doi: 10.1002/mrm.20527.

引用本文的文献

1
Prospective GIRF-based RF phase cycling to reduce eddy current-induced steady-state disruption in bSSFP imaging.
Magn Reson Med. 2020 Jul;84(1):115-127. doi: 10.1002/mrm.28097. Epub 2019 Nov 22.
3
Pseudo-projection-driven, self-gated cardiac cine imaging using cartesian golden step phase encoding.
Magn Reson Med. 2016 Aug;76(2):417-29. doi: 10.1002/mrm.25834. Epub 2015 Oct 31.
5
Whole heart coronary imaging with flexible acquisition window and trigger delay.
PLoS One. 2015 Feb 26;10(2):e0112020. doi: 10.1371/journal.pone.0112020. eCollection 2015.

本文引用的文献

2
Flow compensation in balanced SSFP sequences.
Magn Reson Med. 2005 Oct;54(4):901-7. doi: 10.1002/mrm.20619.
3
Analysis and compensation of eddy currents in balanced SSFP.
Magn Reson Med. 2005 Jul;54(1):129-37. doi: 10.1002/mrm.20527.
6
9
Flow artifacts in steady-state free precession cine imaging.
Magn Reson Med. 2004 Jan;51(1):115-22. doi: 10.1002/mrm.10665.
10
Whole-heart steady-state free precession coronary artery magnetic resonance angiography.
Magn Reson Med. 2003 Dec;50(6):1223-8. doi: 10.1002/mrm.10653.