Suppr超能文献

生物电位记录中表面电极的电极材料长期特性研究。

Long-term characterization of electrode materials for surface electrodes in biopotential recording.

作者信息

Hoffmann Klaus-Peter, Ruff Roman, Poppendieck Wigand

机构信息

Department of Medical Technology and Neural Prosthetics, Fraunhofer IBMT, St. Ingbert, Germany.

出版信息

Conf Proc IEEE Eng Med Biol Soc. 2006;2006:2239-42. doi: 10.1109/IEMBS.2006.260443.

Abstract

The long-term electrical behavior of 16 different electrode materials was investigated by using continuous impedance spectroscopy over a period of 10 days. The materials included bare and electrolytically treated metals, metals coated with intrinsically conductive polymers, and polymers with conductive particles. Electrolytic treatment of metal electrodes yielded a significant impedance reduction. The lowest impedance values could be reached with the polymer-coated metal electrodes. The impedance behavior is an important aspect when choosing an electrode material, and has to be considered in signal processing. The choice of a material depends on the application of the electrode. Moreover, for long-term applications, the electrochemical stability of a material has to be considered.

摘要

通过连续阻抗谱在10天的时间内研究了16种不同电极材料的长期电学行为。这些材料包括裸金属和经过电解处理的金属、涂有本征导电聚合物的金属以及含有导电颗粒的聚合物。对金属电极进行电解处理可显著降低阻抗。聚合物涂层金属电极可达到最低的阻抗值。在选择电极材料时,阻抗行为是一个重要方面,并且在信号处理中必须予以考虑。材料的选择取决于电极的应用。此外,对于长期应用,必须考虑材料的电化学稳定性。