Cai Sheng-Suan, Short Luke Chandler, Syage Jack A, Potvin Michael, Curtis Jonathan M
Syagen Technology, Inc., 1411 Warner Avenue, Tustin, CA 92780, USA.
J Chromatogr A. 2007 Nov 30;1173(1-2):88-97. doi: 10.1016/j.chroma.2007.10.008. Epub 2007 Oct 10.
In this work, we evaluate the performance of liquid chromatography-atmospheric pressure photoionization-mass spectrometry (LC-APPI-MS) for non-aqueous reversed phase analysis of six triacylglycerol model compounds using six binary mobile phases including MeOH/iPrOH, MeOH/CHCl(3), MeOH/CH(2)Cl(2), CH(3)CN/iPrOH, CH(3)CN/CHCl(3), and CH(3)CN/CH(2)Cl(2). All mobile phases give comparably good separation performance on a Gemini C(18) column with carefully adjusted gradient elution programs. APPI sensitivity varies from one mobile phase to the other without dopants; however use of dopants brings sensitivity to comparable levels for all mobile phases. MeOH/iPrOH offers high sensitivity without dopants due to self-doping effect and dopants are not necessary for this mobile phase. Dopants enhance analyte sensitivity to a varying degree for each of the mobile phases tested. Photo-induced chemical ionization (PCI) of solvent may play a significant role in achieving high sensitivity. Two critical parameters affecting sensitivity are photoabsorption cross-sections and ionization potentials of mobile phase solvents. How these mobile phase solvents affect APPI sensitivity and their dependency on dopant use are discussed. All six mobile phases offer comparable overall limits of detection for the analytes tested. These results indicate that LC-APPI-MS is a successful tool for neutral lipid analysis, giving high sensitivity with a variety of non-aqueous mobile phases.
在本研究中,我们使用包括甲醇/异丙醇、甲醇/三氯甲烷、甲醇/二氯甲烷、乙腈/异丙醇、乙腈/三氯甲烷和乙腈/二氯甲烷在内的六种二元流动相,评估了液相色谱 - 大气压光电离 - 质谱联用仪(LC - APPI - MS)对六种三酰甘油模型化合物进行非水反相分析的性能。在经过精心调整梯度洗脱程序的Gemini C18柱上,所有流动相均具有相当良好的分离性能。在不使用掺杂剂的情况下,不同流动相的APPI灵敏度有所差异;然而,使用掺杂剂可使所有流动相的灵敏度达到相当的水平。甲醇/异丙醇由于自掺杂效应,在不使用掺杂剂时就具有高灵敏度,因此该流动相无需使用掺杂剂。对于所测试的每种流动相,掺杂剂均能不同程度地提高分析物的灵敏度。溶剂的光致化学电离(PCI)可能在实现高灵敏度方面发挥重要作用。影响灵敏度的两个关键参数是流动相溶剂的光吸收截面和电离电位。本文讨论了这些流动相溶剂如何影响APPI灵敏度以及它们对掺杂剂使用的依赖性。所有六种流动相对所测试的分析物均具有相当的总体检测限。这些结果表明,LC - APPI - MS是一种用于中性脂质分析的成功工具,能在多种非水流动相条件下实现高灵敏度检测。