电脑化对数最小分辨角视力测量系统(COMPlog)的验证:与成人及弱视儿童的ETDRS和电子ETDRS测试算法的比较

Validation of a computerised logMAR visual acuity measurement system (COMPlog): comparison with ETDRS and the electronic ETDRS testing algorithm in adults and amblyopic children.

作者信息

Laidlaw D A H, Tailor V, Shah N, Atamian S, Harcourt C

机构信息

Consultant Vitreo Retinal Surgeon, St Thomas' Hospital, Lambeth Palace Rd, London SE1 7EH, UK.

出版信息

Br J Ophthalmol. 2008 Feb;92(2):241-4. doi: 10.1136/bjo.2007.121715. Epub 2007 Nov 9.

Abstract

BACKGROUND/AIM: The COMPlog clinical visual acuity measuring system is being developed for both routine and research use. This study aimed to validate its performance in amblyopic children and both normal and diseased adults against the gold standard ETDRS chart and the E-ETDRS computerised acuity measurement algorithm.

METHOD

Timed test and retest fully interpolated five letters per line logMAR visual acuity measurements were taken for 70 adults and 59 amblyopic children using the ETDRS chart and the COMPlog visual acuity measurement system. 39 of the adults also underwent computerised acuity testing using the E-ETDRS testing algorithm. The tested adults included normals as well as subjects with a range of ocular diseases. The methods of Bland and Altman were employed with test-retest variability (TRV) expressed as 95% confidence limits for agreement.

RESULTS

No significant bias was observed between the gold standard ETDRS acuity measurements and those taken with either COMPlog or E-ETDRS. TRVs of +/-0.12 logMAR and +/-0.10 logMAR were respectively found for COMPlog measurements in the amblyopic children and adult groups compared with +/-0.12 logMAR for the ETDRS chart in both groups. The TRV of the E-ETDRS system was slightly greater at +/-0.16 logMAR. Median testing times for COMPlog and ETDRS were 95 and 85 s and 66 and 56 s respectively in the paediatric and adult groups and 120 s for the E-ETDRS measurements on adults.

DISCUSSION

COMPlog measurements agree well with and are similarly reliable to the gold standard ETDRS chart with comparable test times. E-ETDRS algorithm measurements took approximately twice as long.

摘要

背景/目的:COMPlog临床视力测量系统正在开发中,用于常规和研究用途。本研究旨在针对金标准ETDRS视力表以及E-ETDRS计算机化视力测量算法,验证其在弱视儿童以及正常和患病成年人中的性能。

方法

使用ETDRS视力表和COMPlog视力测量系统,对70名成年人和59名弱视儿童进行了每行完全插值五个字母的定时测试和复测logMAR视力测量。39名成年人还使用E-ETDRS测试算法进行了计算机化视力测试。受试成年人包括正常人和患有一系列眼部疾病的受试者。采用Bland和Altman方法,将重测变异性(TRV)表示为一致性的95%置信限。

结果

在金标准ETDRS视力测量与COMPlog或E-ETDRS测量之间未观察到显著偏差。与两组中ETDRS视力表的±0.12 logMAR相比,弱视儿童组和成人组中COMPlog测量的TRV分别为±0.12 logMAR和±0.10 logMAR。E-ETDRS系统的TRV略大,为±0.16 logMAR。儿科组和成人组中COMPlog和ETDRS的中位测试时间分别为95秒和85秒,以及66秒和56秒,成人的E-ETDRS测量为120秒。

讨论

COMPlog测量与金标准ETDRS视力表一致性良好,可靠性相似,测试时间相当。E-ETDRS算法测量时间大约是其两倍。