Suppr超能文献

评估一种旨在改善基于人群调查中视力评估的新型logMAR视力表。

Evaluating a new logMAR chart designed to improve visual acuity assessment in population-based surveys.

作者信息

Bourne R R A, Rosser D A, Sukudom P, Dineen B, Laidlaw D A H, Johnson G J, Murdoch I E

机构信息

Department of Epidemiology and International Eye Health Institute of Ophthalmology University College London UK.

出版信息

Eye (Lond). 2003 Aug;17(6):754-8. doi: 10.1038/sj.eye.6700500.

Abstract

AIMS

This study aimed to evaluate a new chart designed to improve the collection of visual acuity data in population-based surveys. The Reduced logMAR E chart (RLME) employs three letters per line, 'tumbling E' optotypes, and conforms to accepted contemporary design principles.

METHODS

The performance of the chart was assessed within a population-based glaucoma survey in Thailand. Performance indices were test-retest variability (TRV) and agreement with acuity data measured using the ETDRS logMAR chart which acted as the 'gold standard'.

RESULTS

The 95% confidence limits for TRV of RLME acuity data were +/-0.15 logMAR. This figure is consistent with published data on the TRV of acuities measured using five-letter-per-line logMAR charts. The mean difference between RLME and ETDRS acuity data was 0.00 logMAR (95% confidence intervals of +/-0.05 logMAR) indicating that RLME acuities agreed well with those of the ETDRS chart. The chart and its method of use was readily accepted by the local ancillary staff who required only minimal training before acuity measurement could be delegated to them.

CONCLUSIONS

The study demonstrated that the RLME chart is capable of accurate and repeatable acuity measurements. Certain aspects of the design of the RLME chart may be particularly pertinent to the measurement of vision in population-based surveys.

摘要

目的

本研究旨在评估一种新设计的视力表,该视力表旨在改善基于人群的调查中视力数据的收集。简化对数最小分辨角E视力表(RLME)每行有三个字母,采用“翻转E”视标,符合当代公认的设计原则。

方法

在泰国一项基于人群的青光眼调查中评估该视力表的性能。性能指标为重测变异性(TRV)以及与使用ETDRS对数最小分辨角视力表测量的视力数据的一致性,ETDRS对数最小分辨角视力表作为“金标准”。

结果

RLME视力数据TRV的95%置信限为±0.15对数最小分辨角。该数值与已发表的关于使用每行五个字母的对数最小分辨角视力表测量视力的TRV数据一致。RLME与ETDRS视力数据的平均差异为0.00对数最小分辨角(95%置信区间为±0.05对数最小分辨角),表明RLME视力与ETDRS视力表的视力高度一致。当地辅助人员很容易接受该视力表及其使用方法,他们在被委托进行视力测量之前只需要极少的培训。

结论

该研究表明RLME视力表能够进行准确且可重复的视力测量。RLME视力表设计的某些方面可能与基于人群的调查中的视力测量特别相关。