Hughes S M, Blake B L, Woods S L, Lehmann C U
Eudowood Neonatal Pulmonary Division, Department of Pediatrics, The Johns Hopkins University, Baltimore, MD, USA.
J Perinatol. 2007 Dec;27(12):800-1. doi: 10.1038/sj.jp.7211831.
A term infant requiring resuscitation was found to have a false-positive color change on a colorimetric carbon dioxide device as a result of administration of epinephrine via an endotracheal tube. Using models of direct application and vapor exposure with a test lung, we discovered that epinephrine, atropine, infasurf and naloxone may result in false-positive color change. This false-positive response may lead to delayed recognition of esophageal intubation.