Suppr超能文献

3.0T和1.5T下双手的双侧对比增强磁共振血管造影:健康志愿者定量和定性图像参数的个体内比较

Bilateral ce-MR angiography of the hands at 3.0 T and 1.5 T: intraindividual comparison of quantitative and qualitative image parameters in healthy volunteers.

作者信息

Winterer Jan Thorsten, Moske-Eick Olaf, Markl Michael, Frydrychowicz Alexander, Bley Thorsten Alexander, Langer Mathias

机构信息

Department of Diagnostic Radiology, University of Freiburg, Hugstetter Str. 55, 79106, Freiburg, Germany.

出版信息

Eur Radiol. 2008 Apr;18(4):658-64. doi: 10.1007/s00330-007-0800-1. Epub 2007 Nov 27.

Abstract

The purpose of this study was to determine the benefit of bilateral contrast-enhanced MR angiography (ce-MRA) of the hands at 3.0 Tesla (T) compared with an established 1.5-T technique in healthy volunteers. Intraindividual bilateral ce-MRA of the hands was performed at 1.5 T and 3.0 T in 14 healthy volunteers using a timed ultra-fast GRE sequence featuring parallel acquisition. The evaluation comprised measurement of the vessel signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), rating of the image quality and the assessment of artefacts and venous contamination. At 3.0 T, SNR improved up to 95% and CNR up to 129%. The image quality of the larger inflow arteries, the palm arches and common digital arteries was good or sufficient at either magnetic field strengths. However, 3.0-T MRA was clearly superior in the depiction of the digital arteries. Ce-MRA of the hand clearly profits from the use of 3.0 T. Compared with 1.5 T, a substantial increase of CNR is found resulting in a significantly better delineation of the small digital arteries. Saturation affects more the SNR of the perivascular tissue than the contrast-enhanced blood, and thus leads to a marked increase of CNR at 3.0.

摘要

本研究的目的是在健康志愿者中,确定与已确立的1.5T技术相比,3.0特斯拉(T)手部双侧对比增强磁共振血管造影(ce-MRA)的优势。14名健康志愿者在1.5T和3.0T条件下,使用具有并行采集功能的定时超快速梯度回波序列进行手部个体内双侧ce-MRA检查。评估内容包括测量血管的信噪比(SNR)和对比噪声比(CNR)、对图像质量进行评分以及评估伪影和静脉污染情况。在3.0T时,SNR提高了95%,CNR提高了129%。在两种磁场强度下,较大的流入动脉、掌弓和指总动脉的图像质量均良好或足够。然而,3.0T MRA在显示指动脉方面明显更具优势。手部ce-MRA明显受益于3.0T的使用。与1.5T相比,CNR显著增加,从而能更清晰地勾勒出小指动脉。饱和对血管周围组织SNR的影响大于对对比增强血液的影响,因此在3.0T时导致CNR显著增加。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验