一项关于Lifeshirt与标准多导睡眠图在检测阻塞性睡眠呼吸暂停方面有效性的调查。

An investigation of the validity of the Lifeshirt in comparison to standard polysomnography in the detection of obstructive sleep apnea.

作者信息

Goodrich Suanne, Orr William C

机构信息

Lynn Health Science Institute, 5300 N. Independence, Ste 130, Oklahoma City, OK 73112, USA.

出版信息

Sleep Med. 2009 Jan;10(1):118-22. doi: 10.1016/j.sleep.2007.11.003. Epub 2008 Feb 20.

Abstract

BACKGROUND

Due to the cost and waiting times for traditional sleep studies, there is great interest in finding alternatives for the diagnosis of sleep apnea. Several alternatives exist, including ambulatory devices. Our laboratory utilized one such device, a form-fitting vest called the Lifeshirt, and compared its accuracy to that of polysomnography (PSG).

METHODS

PSG was performed with simultaneous utilization of the Lifeshirt on 50 individuals who met screening criteria for obstructive sleep apnea. Participants came to the sleep laboratory approximately 2 h before their normal bedtime. A sleep technician prepared each participant for PSG and Lifeshirt monitoring. "Lights Out" occurred when the participant was ready for bed, and time in bed was standardized to approximately 7 h. PSGs were scored by experienced personnel in our laboratory, while Lifeshirt data were sent electronically to the Lifeshirt manufacturer for analysis. The major variable that was compared between PSG and Lifeshirt was the apnea hypopnea index (AHI), or the number of apneas and hypopneas per hour. Due to incomplete data on two participants, analyses were completed on 48 individuals.

RESULTS

Sensitivity of the Lifeshirt ranged from .85 (AHI of > or =5) to 1.00 (AHI of > or =25). Specificity ranged from .67 to 1.00. Using the Bland-Altman technique of determining agreement, the mean difference between the AHI of the Lifeshirt and PSG was 1.02 (+/-16.36). When these values are plotted, every case falls within the limits of agreement, with one exception.

CONCLUSIONS

The Lifeshirt compared favorably with PSG and could be used with considerable confidence for the screening of patients with suspected obstructive sleep apnea.

摘要

背景

由于传统睡眠研究的成本和等待时间,人们对寻找睡眠呼吸暂停诊断的替代方法有着浓厚的兴趣。存在几种替代方法,包括便携式设备。我们的实验室使用了一种这样的设备,一种名为“生命衬衫”的贴合身体的背心,并将其准确性与多导睡眠图(PSG)进行了比较。

方法

对50名符合阻塞性睡眠呼吸暂停筛查标准的个体同时使用PSG和“生命衬衫”。参与者在正常就寝时间前约2小时来到睡眠实验室。一名睡眠技术人员为每位参与者准备PSG和“生命衬衫”监测。当参与者准备睡觉时“熄灯”,卧床时间标准化为约7小时。PSG由我们实验室的经验丰富的人员进行评分,而“生命衬衫”数据则通过电子方式发送给“生命衬衫”制造商进行分析。PSG和“生命衬衫”之间比较的主要变量是呼吸暂停低通气指数(AHI),即每小时呼吸暂停和低通气的次数。由于两名参与者的数据不完整,对48名个体进行了分析。

结果

“生命衬衫”的敏感性范围为0.85(AHI≥5)至1.00(AHI≥25)。特异性范围为0.67至1.00。使用Bland-Altman确定一致性的技术,“生命衬衫”和PSG的AHI之间的平均差异为1.02(±16.36)。当绘制这些值时,除了一个例外,每个病例都落在一致性限度内。

结论

“生命衬衫”与PSG相比具有优势,可相当有信心地用于筛查疑似阻塞性睡眠呼吸暂停的患者。