种植体植入前计算机断层扫描评估下颌部位的准确性。

Accuracy of computerized tomography for the evaluation of mandibular sites prior to implant placement.

作者信息

Sforza Nicola Marco, Franchini Federico, Lamma Achille, Botticelli Susanna, Ghigi Gino

机构信息

Studio Associato Sforza, Piazza Aldrovandi 12, Bologna, Italy.

出版信息

Int J Periodontics Restorative Dent. 2007 Dec;27(6):589-95.

Abstract

The objective of the present study was to observe the effect of positioning of the mandible on the accuracy of cross-sectional images obtained by reformatting computerized tomographic (CT) scans. An additional aim was to evaluate the ability of a software program (DentalVox, Era Scientific) to reconstruct these measurements on the reformatted images, regardless of the positioning of the mandible, accurately and without distortion. The test was carried out by examining a partially edentulous dry human mandible with an acrylic radiologic template. Through the use of an acrylic glass support, the mandible was positioned at angles of 0, 10, 15, 20, and 30 degrees relative to the scanning gantry, and a series of CT scans was performed that provided five sets of axial images. Each set of original axial images was reformatted by the DentalVox software, used first in its basic function, which is typical of all software for axial CT measurement (control group), and again in its function of site-specific multiplanar reconstruction (test group). The results showed that the position of the mandible in relation to the CT gantry can influence the precision of the linear measurements. The error ranged from 2% to 51%. The DentalVox software allowed the reconstruction of cross-sectional images with very little distortion regardless of the mandibular position.

摘要

本研究的目的是观察下颌骨的位置对通过计算机断层扫描(CT)扫描重新格式化获得的横断面图像准确性的影响。另一个目的是评估一个软件程序(DentalVox,Era Scientific)在重新格式化图像上重建这些测量值的能力,无论下颌骨的位置如何,都能准确且无失真地进行。该测试通过使用带有丙烯酸放射模板的部分无牙干燥人类下颌骨进行。通过使用丙烯酸玻璃支架,将下颌骨相对于扫描机架以0、10、15、20和30度的角度定位,并进行一系列CT扫描,提供五组轴向图像。每组原始轴向图像首先通过DentalVox软件以其基本功能进行重新格式化,这是所有轴向CT测量软件的典型功能(对照组),然后再次以其特定部位多平面重建功能进行重新格式化(测试组)。结果表明,下颌骨相对于CT机架的位置会影响线性测量的精度。误差范围为2%至51%。无论下颌位置如何,DentalVox软件都能以极小的失真重建横断面图像。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验