Method of Individual Adjustment for 3D CT Analysis: Linear Measurement.

作者信息

Kim Dong Kyu, Choi Dong Hun, Lee Jeong Woo, Yang Jung Dug, Chung Ho Yun, Cho Byung Chae, Choi Kang Young

机构信息

Department of Plastic and Reconstructive Surgery, School of Medicine, Kyungpook National University, Daegu, Republic of Korea.

出版信息

Biomed Res Int. 2016;2016:6893072. doi: 10.1155/2016/6893072. Epub 2016 Dec 13.

Abstract

. We aim to regularize measurement values in three-dimensional (3D) computed tomography (CT) reconstructed images for higher-precision 3D analysis, focusing on length-based 3D cephalometric examinations. . We measure the linear distances between points on different skull models using Vernier calipers (real values). We use 10 differently tilted CT scans for 3D CT reconstruction of the models and measure the same linear distances from the picture archiving and communication system (PACS). In both cases, each measurement is performed three times by three doctors, yielding nine measurements. The real values are compared with the PACS values. Each PACS measurement is revised based on the display field of view (DFOV) values and compared with the real values. . The real values and the PACS measurement changes according to tilt value have no significant correlations ( > 0.05). However, significant correlations appear between the real values and DFOV-adjusted PACS measurements ( < 0.001). Hence, we obtain a correlation expression that can yield real physical values from PACS measurements. The DFOV value intervals for various age groups are also verified. . Precise confirmation of individual preoperative length and precise analysis of postoperative improvements through 3D analysis is possible, which is helpful for facial-bone-surgery symmetry correction.

摘要

我们旨在对三维(3D)计算机断层扫描(CT)重建图像中的测量值进行规范化处理,以实现更高精度的3D分析,重点关注基于长度的3D头影测量检查。我们使用游标卡尺(实际值)测量不同颅骨模型上各点之间的线性距离。我们使用10次不同倾斜角度的CT扫描对模型进行3D CT重建,并从图像存档与通信系统(PACS)中测量相同的线性距离。在这两种情况下,每项测量均由三位医生进行三次,共得到九次测量结果。将实际值与PACS值进行比较。每次PACS测量均根据显示视野(DFOV)值进行修正,并与实际值进行比较。实际值与根据倾斜值变化的PACS测量值之间无显著相关性(>0.05)。然而,实际值与经DFOV调整后的PACS测量值之间存在显著相关性(<0.001)。因此,我们得到了一个可以从PACS测量值中得出实际物理值的相关表达式。还对不同年龄组的DFOV值区间进行了验证。通过3D分析可以精确确认个体术前长度,并对术后改善情况进行精确分析,这有助于面部骨骼手术的对称矫正。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/f3b7/5187594/dd9105bb2949/BMRI2016-6893072.001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验