准分子激光原位角膜磨镶术(LASIK)前后通过非接触式角膜内皮显微镜和超声角膜测厚仪测量的中央角膜厚度比较。

Comparison of central corneal thickness as measured by non-contact specular microscopy and ultrasound pachymetry before and post LASIK.

作者信息

Zhao Ming-hui, Zou Jun, Wang Wen-qing, Li Jun

机构信息

Department of Ophthalmology, Shanghai Jiaotong University Affiliated Sixth People's Hospital, Shanghai, China.

出版信息

Clin Exp Ophthalmol. 2007 Dec;35(9):818-23. doi: 10.1111/j.1442-9071.2007.01633.x.

Abstract

PURPOSE

To compare central corneal thickness (CCT) measurements made by ultrasound pachymetry and non-contact specular microscopy in eyes before and post laser in situ keratomileusis (LASIK).

METHODS

Forty eyes of 20 consecutive refractive surgery candidates were included in this study. The CCT in both eyes was measured before and 3 months after LASIK with a non-contact specular microscope and a ultrasound pachymeter.

RESULT

Both before and after LASIK, the difference of the mean CCT between non-contact specular microscopy and ultrasound pachymetry was statistically significant (P < 0.001 before surgery, and P = 0.02 after surgery). The difference between them was -14.41 microm and -3.69 microm, respectively. In both situations, non-contact specular microscopy and ultrasound pachymetry measurements were highly correlated (r = 0.959, before surgery, and r = 0.979 after LASIK). Bland-Altman analysis showed that either before or post LASIK, the 95% limits of agreement have a span of approximately 30 microm. Preoperatively, the mean standard deviation of repeated measurements was 4.26 microm by ultrasound pachymetry, and 4.02 microm by non-contact specular microscopy, whereas it was 3.83 microm by ultrasound pachymetry, and 3.89 microm by non-contact specular microscopy, postoperatively. No statistically significant differences were found between the mean standard deviations by these two devices in both situations (P = 0.561 before surgery, and P = 0.849 after LASIK).

CONCLUSIONS

Both before and after LASIK, non-contact specular microscopy is not in complete agreement with ultrasound pachymetry in measuring CCT. Both devices provided comparable repeatability of measurements.

摘要

目的

比较准分子原位角膜磨镶术(LASIK)术前和术后使用超声角膜测厚仪和非接触式角膜内皮显微镜测量中央角膜厚度(CCT)的结果。

方法

本研究纳入了20例连续的屈光手术候选者的40只眼睛。使用非接触式角膜内皮显微镜和超声角膜测厚仪在LASIK术前和术后3个月测量双眼的CCT。

结果

LASIK术前和术后,非接触式角膜内皮显微镜与超声角膜测厚仪测量的平均CCT差异均具有统计学意义(术前P < 0.001,术后P = 0.02)。两者差异分别为-14.41微米和-3.69微米。在两种情况下,非接触式角膜内皮显微镜和超声角膜测厚仪的测量结果均高度相关(LASIK术前r = 0.959,术后r = 0.979)。Bland-Altman分析显示,LASIK术前或术后,95%一致性界限的范围约为30微米。术前,超声角膜测厚仪重复测量的平均标准差为4.26微米,非接触式角膜内皮显微镜为4.02微米;术后,超声角膜测厚仪为3.83微米,非接触式角膜内皮显微镜为3.89微米。两种设备在两种情况下的平均标准差之间均未发现统计学显著差异(术前P = 0.561,术后P = 0.849)。

结论

LASIK术前和术后,非接触式角膜内皮显微镜在测量CCT方面与超声角膜测厚仪不完全一致。两种设备提供了相当的测量重复性。