Al-Ageel Saleh, Al-Muammar Abdulrahman M
Department of Ophthalmology, College of Medicine, King Saud University, Riyadh, Saudi Arabia.
Saudi J Ophthalmol. 2009 Oct;23(3-4):181-7. doi: 10.1016/j.sjopt.2009.10.002. Epub 2009 Oct 24.
To compare central corneal thickness (CCT) measurements taken with Pentacam, noncontact specular microscope (NCSM), and ultrasound pachymetry (US) in normal and post-laser in situ keratomileusis (LASIK) eyes and to assess the agreement between the three devices.
Prospective clinical trial.
Central corneal thickness (CCT) was measured in 94 eyes of normal volunteer and in 72 eyes of post-LASIK patients. Measurements were made with the three devices.
In normal eyes, the mean (±SD) CCT taken with Pentacam, NCSM, and US was 552.6 ± 36.8 μm, 511.9 ± 38.6 μm, and 533.3 ± 37.9 μm, respectively. The average values of CCT taken with the three instruments were significantly different. In post-LASIK eyes the mean CCT with Pentacam, NCSM, and US was 483.02 ± 6.03 μm, 450.7 ± 5.3 μm, and 469.5 ± 5.8 μm, respectively. The average values of CCT taken were significantly different for Pentacam vs. NCSM (P = 0.046) and Pentacam vs. US (P = 0.02), but not significant for NCSM vs. US (P = 0.352). The Bland and Altman method for assessing clinical agreement between 3 instruments showed that in normal eyes, the mean values and paired differences of the three CCT devices were found to be statistically independent. In post-LASIK eyes, there was significant association between the difference and the mean of the Pentacam and NCSM, and US and NCSM.
The three devices cannot be used interchangeably in normal and post-LASIK eyes. Pentacam tends to give significantly thicker reading than ultrasound pachymetry.
比较Pentacam、非接触式角膜内皮显微镜(NCSM)和超声角膜测厚仪(US)在正常眼和准分子原位角膜磨镶术(LASIK)术后眼中测量中央角膜厚度(CCT)的结果,并评估这三种设备之间的一致性。
前瞻性临床试验。
对94只正常志愿者的眼睛和72只LASIK术后患者的眼睛进行中央角膜厚度(CCT)测量。使用这三种设备进行测量。
在正常眼中,Pentacam、NCSM和US测量的平均(±标准差)CCT分别为552.6±36.8μm、511.9±38.6μm和533.3±37.9μm。三种仪器测量的CCT平均值有显著差异。在LASIK术后眼中,Pentacam、NCSM和US测量的平均CCT分别为483.02±6.03μm、450.7±5.3μm和469.5±5.8μm。Pentacam与NCSM(P = 0.046)以及Pentacam与US(P = 0.02)测量的CCT平均值差异有统计学意义,但NCSM与US(P = 0.352)差异无统计学意义。用于评估三种仪器临床一致性的Bland和Altman方法显示,在正常眼中,三种CCT设备的平均值和配对差异在统计学上是独立的。在LASIK术后眼中,Pentacam与NCSM、US与NCSM的差异与平均值之间存在显著相关性。
在正常眼和LASIK术后眼中,这三种设备不能互换使用。Pentacam测量的结果往往比超声角膜测厚仪测量的结果明显更厚。