正常眼及飞秒激光原位角膜磨镶术后眼的中央角膜厚度的光学低相干反射测量法与超声测厚法测量

Measurement of central corneal thickness with optical low-coherence reflectometry and ultrasound pachymetry in normal and post-femtosecond laser in situ keratomileusis eyes.

作者信息

Huang Jinhai, Liao Na, Savini Giacomo, Li Yuanguang, Bao Fangjun, Yu Ye, Yu Ayong, Wang Qinmei

机构信息

*Eye Hospital, Ophthalmology Institute, School of Optometry and Ophthalmology, Wenzhou Medical University, Wenzhou, China; †Key Laboratory of Vision Science, Ophthalmology Institute, Ministry of Health P.R. China, Wenzhou, China; and ‡G.B. Bietti Foundation IRCCS, Rome, Italy.

出版信息

Cornea. 2015 Feb;34(2):204-8. doi: 10.1097/ICO.0000000000000329.

Abstract

PURPOSE

To determine the repeatability and reproducibility of measurements of central corneal thickness (CCT) using optical low-coherence reflectometry (Lenstar LS900; Haag Streit) in normal eyes and post-femtosecond laser in situ keratomileusis (post-FS-LASIK) eyes and evaluate their agreement with ultrasound (US) pachymetry.

METHODS

CCT was measured using Lenstar and US pachymetry sequentially in normal and post-FS-LASIK eyes by 2 experienced observers. Intraoperator repeatability and interoperator reproducibility were assessed by within-subject standard deviation, test-retest repeatability, coefficient of variation (CoV), and intraclass correlation coefficient. Paired t-tests and Bland-Altman plots were used for analyzing agreement between the 2 devices.

RESULTS

In this study, 55 healthy subjects and 50 post-FS-LASIK patients were recruited. Test-retest repeatability of Lenstar was within 10 μm, CoV was less than 1.0%, and intraclass correlation coefficient was more than 0.9 in both normal and post-FS-LASIK groups. Mean difference between these methods was 1.4 ± 4.2 μm and -1.7 ± 5.4 μm, respectively. Moreover, measurements of CCT showed narrow 95% limits of agreement (range, normal group: -6.8 and 9.6 μm; post-FS-LASIK group: -12.4 and 8.9 μm), which implied good agreement.

CONCLUSIONS

Measurements of CCT using Lenstar showed excellent intraoperator repeatability and interoperator reproducibility both in normal eyes and post-FS-LASIK eyes. Measurements of CCT using Lenstar and US pachymetry showed good agreement and both can be used interchangeably.

摘要

目的

使用光学低相干反射仪(Lenstar LS900;Haag Streit)测定正常眼和飞秒激光原位角膜磨镶术(FS-LASIK)术后眼中央角膜厚度(CCT)测量的重复性和再现性,并评估其与超声(US)测厚法的一致性。

方法

由2名经验丰富的观察者对正常眼和FS-LASIK术后眼依次使用Lenstar和US测厚法测量CCT。通过受试者内标准差、重测重复性、变异系数(CoV)和组内相关系数评估操作者内重复性和操作者间再现性。使用配对t检验和Bland-Altman图分析两种设备之间的一致性。

结果

本研究招募了55名健康受试者和50名FS-LASIK术后患者。在正常组和FS-LASIK术后组中,Lenstar的重测重复性在10μm以内,CoV小于1.0%,组内相关系数大于0.9。这些方法之间的平均差异分别为1.4±4.2μm和-1.7±5.4μm。此外,CCT测量显示一致性的95%界限较窄(范围,正常组:-6.8至9.6μm;FS-LASIK术后组:-12.4至8.9μm),这意味着一致性良好。

结论

使用Lenstar测量CCT在正常眼和FS-LASIK术后眼中均显示出优异的操作者内重复性和操作者间再现性。使用Lenstar和US测厚法测量CCT显示出良好的一致性,两者可互换使用。