PURPOSE

To assess the three-dimensional repeatability of thickness measurements for epithelium, stroma, cornea, flap, and residual stromal bed using the Artemis very high-frequency (VHF) digital ultrasound arc-scanner (ArcScan Inc).

METHODS

Five consecutive measurements were obtained for 10 eyes of 10 patients 1 year after LASIK using the Artemis VHF digital ultrasound arc-scanner across the central 10-mm diameter of the cornea. Repeatability analysis was performed for thickness measurements for each corneal layer-epithelium, stroma, cornea, flap, and residual stromal bed. The standard deviation of repeated measurements (point-repeatability) was calculated for each measurement location in 0.1-mm steps for the 10×10-mm matrix. The pooled standard deviation of the point-repeatability for each measurement location within the central 1-, 2-, and 3-mm radius was calculated (region-repeatability). The corneal thickness of the baseline scan set was compared to that of subsequent scan sets within the same session and plotted over time to assess any possible hydration effects of the immersion technique.

RESULTS

The repeatability at the corneal vertex was 0.58 μm for epithelium, 1.78 μm for stroma, 1.68 μm for cornea, 1.68 μm for flap, and 2.27 μm for residual stromal bed. The region-repeatability within the central 1-mm radius was 1.01 μm for epithelium, 3.44 μm for stroma, 3.35 μm for cornea, 2.81 μm for flap, and 3.97 μm for residual stromal bed. The mean difference in corneal thickness from the baseline value was within 1.25 μm for each of the subsequent four scan sets over a 5-minute immersion period.

CONCLUSIONS

Layered pachymetry of the epithelium, stroma, cornea, flap, and residual stromal bed showed high repeatability with the Artemis VHF digital ultrasound arc-scanner. The high repeatability validates the use of the Artemis for in vivo layered pachymetry.

摘要

目的

使用 Artemis 甚高频(VHF)数字超声弧形扫描仪评估上皮、基质、角膜、瓣和残余基质床的厚度测量的三维重复性。

方法

对 10 例(10 只眼)LASIK 术后 1 年的患者,使用 Artemis VHF 数字超声弧形扫描仪在角膜中央 10mm 直径范围内连续测量 5 次。对每个角膜层-上皮、基质、角膜、瓣和残余基质床的厚度测量进行重复性分析。以 0.1mm 为步长,在 10×10mm 矩阵的每个测量位置计算重复测量的标准差(点重复性)。计算中央 1、2 和 3mm 半径内每个测量位置的点重复性的汇总标准差(区域重复性)。将基线扫描设置的角膜厚度与同一时段内后续扫描设置进行比较,并随时间绘制,以评估浸液技术的任何可能的水化效应。

结果

角膜顶点的重复性为上皮 0.58μm、基质 1.78μm、角膜 1.68μm、瓣 1.68μm 和残余基质床 2.27μm。中央 1mm 半径内的区域重复性为上皮 1.01μm、基质 3.44μm、角膜 3.35μm、瓣 2.81μm 和残余基质床 3.97μm。在 5 分钟的浸液过程中,从基线值开始,每个后续的四个扫描设置的角膜厚度平均差值均在 1.25μm 以内。

结论

上皮、基质、角膜、瓣和残余基质床的分层测厚显示出 Artemis VHF 数字超声弧形扫描仪的高度重复性。这种高度的可重复性验证了 Artemis 用于活体分层测厚的用途。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0c18/4464782/e0a475b5d17f/nihms695884f1.jpg

相似文献

引用本文的文献

2
Corneal Epithelial Thickness Mapping: A Major Review.角膜上皮厚度地形图:一项主要综述。
J Ophthalmol. 2024 Jan 2;2024:6674747. doi: 10.1155/2024/6674747. eCollection 2024.
3
Applications of epithelial thickness mapping in corneal refractive surgery.上皮厚度测绘在角膜屈光手术中的应用。
Saudi J Ophthalmol. 2022 Jul 11;36(1):25-35. doi: 10.4103/sjopt.sjopt_227_21. eCollection 2022 Jan-Mar.
10
Early diagnosis of keratoconus: what difference is it making?圆锥角膜的早期诊断:它有何作用?
Br J Ophthalmol. 2014 Nov;98(11):1465-6. doi: 10.1136/bjophthalmol-2014-305120. Epub 2014 Apr 23.

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验