Grimaldi E, Del Vecchio L, Scopacasa F, Lo Pardo C, Capone F, Pariante S, Scalia G, De Caterina M
Department of Biochemistry and Medical Biotechnology, University Federico II, Naples, Italy.
Int J Lab Hematol. 2009 Apr;31(2):151-60. doi: 10.1111/j.1751-553X.2007.01015.x. Epub 2007 Dec 20.
The Abbot Cell-Dyn Sapphire is a new generation haematology analyser. The system uses optical/fluorescence flow cytometry in combination with electronic impedance to produce a full blood count. Optical and impedance are the default methods for platelet counting while automated CD61-immunoplatelet analysis can be run as selectable test. The aim of this study was to determine the platelet count performance of the three counting methods available on the instrument and to compare the results with those provided by Becton Dickinson FACSCalibur flow cytometer used as reference method. A lipid interference experiment was also performed. Linearity, carryover and precision were good, and satisfactory agreement with reference method was found for the impedance, optical and CD61-immunoplatelet analysis, although this latter provided the closest results in comparison with flow cytometry. In the lipid interference experiment, a moderate inaccuracy of optical and immunoplatelet counts was observed starting from a very high lipid value.
雅培Cell-Dyn Sapphire是一款新一代血液学分析仪。该系统采用光学/荧光流式细胞术结合电阻抗技术进行全血细胞计数。光学法和电阻抗法是血小板计数的默认方法,而自动CD61免疫血小板分析可作为可选测试项目运行。本研究的目的是确定该仪器上三种可用计数方法的血小板计数性能,并将结果与用作参考方法的贝克曼库尔特FACSCalibur流式细胞仪提供的结果进行比较。还进行了脂质干扰实验。线性、携带污染和精密度良好,电阻抗法、光学法和CD61免疫血小板分析与参考方法的结果一致性令人满意,尽管与流式细胞术相比,后者提供的结果最接近。在脂质干扰实验中,从非常高的脂质值开始,观察到光学法和免疫血小板计数存在中度不准确。