Kenigsberg David N, Mirchandani Sunil, Dover Amanda N, Kowalski Marcin, Wood Mark A, Shepard Richard K, Kalahasty Gautham, Stein Kenneth M, Markowitz Steven M, Iwai Sei, Shah Bindi K, Lerman Bruce B, Mittal Suneet, Ellenbogen Kenneth A
Virginia Commonwealth University Medical Center, Richmond, Virginia 23298-0053, USA.
J Cardiovasc Electrophysiol. 2008 Mar;19(3):270-4. doi: 10.1111/j.1540-8167.2007.01058.x. Epub 2007 Dec 20.
The diameter of implantable cardioverter-defibrillator (ICD) leads has become progressively smaller over time. However, the long-term performance characteristics of these smaller ICD leads are unknown.
We retrospectively evaluated 357 patients who underwent implantation of a Medtronic Sprint Fidelis defibrillating lead at two separate centers between September 2004 and October 2006. Lead characteristics were measured at implant, at early follow-up (1-4 days post implant), and every 3-6 months thereafter.
During the study period, 357 patients underwent implantation of the Medtronic Sprint Fidelis lead. The mean R-wave measured at implant through the device was not different (P = NS) when compared with that measured at first follow-up (10.5 +/- 5.0 mV vs 10.7 +/- 5.1 mV). Forty-one patients (13%) had an R-wave amplitude <or= 5 mV measured through the device at implant. Of those patients with an R-wave amplitude <or= 5 mV at implant measured through the device, 63% (n = 26) remained <or= 5 mV for the duration of follow-up. The mean time to R-wave amplitude <or= 5 mV was 96.2 +/- 123 days. During follow-up, 65 (18%) patients developed R-wave <or= 5 mV. Overall 10 lead revisions (2.8%) were performed during the first year of follow-up.
Abnormal R-wave sensing is frequently observed during follow-up with the Medtronic Fidelis ICD lead. Lead revision was necessary in 2.8% of the patients, most often (8 of 10) due to abnormal R-wave sensing along with elevated pacing threshold. Whether this issue is limited to this lead or reflects a potential problem with all downsized ICD leads merits further investigation.
随着时间的推移,植入式心脏复律除颤器(ICD)导线的直径已逐渐变小。然而,这些较细ICD导线的长期性能特征尚不清楚。
我们回顾性评估了2004年9月至2006年10月期间在两个独立中心植入美敦力Sprint Fidelis除颤导线的357例患者。在植入时、早期随访(植入后1 - 4天)以及此后每3 - 6个月测量导线特征。
在研究期间,357例患者植入了美敦力Sprint Fidelis导线。通过设备在植入时测量的平均R波与首次随访时测量的相比无差异(P = 无显著性差异)(10.5±5.0 mV对10.7±5.1 mV)。41例患者(13%)在植入时通过设备测量的R波振幅≤5 mV。在那些植入时通过设备测量R波振幅≤5 mV的患者中,63%(n = 26)在随访期间一直≤5 mV。R波振幅≤5 mV的平均时间为96.2±123天。在随访期间,65例(18%)患者出现R波≤5 mV。在随访的第一年,共进行了10次导线修正(2.8%)。
在随访美敦力Fidelis ICD导线时经常观察到异常R波感知。2.8%的患者需要进行导线修正,最常见的原因(10例中的8例)是异常R波感知以及起搏阈值升高。这个问题是仅限于这种导线还是反映了所有小型化ICD导线的潜在问题,值得进一步研究。