Keil Adam, Hernandez-Soto Heriberto, Noll Robert J, Fico Miriam, Gao Liang, Ouyang Zheng, Cooks R Graham
Chemistry Department, Purdue University, 560 Oval Drive, West Lafayette, Indiana 47907, USA.
Anal Chem. 2008 Feb 1;80(3):734-41. doi: 10.1021/ac070906o. Epub 2008 Jan 9.
A miniature, handheld mass spectrometer, based on the rectilinear ion trap mass analyzer, has been applied to air monitoring for traces of toxic compounds. The instrument is battery-operated, hand-portable, and rugged. We anticipate its use in public safety, industrial hygiene, and environmental monitoring. Gaseous samples of nine toxic industrial compounds, phosgene, ethylene oxide, sulfur dioxide, acrylonitrile, cyanogen chloride, hydrogen cyanide, acrolein, formaldehyde, and ethyl parathion, were tested. A sorption trap inlet was constructed to serve as the interface between atmosphere and the vacuum chamber of the mass spectrometer. After selective collection of analytes on the sorbent bed, the sorbent tube was evacuated and then heated to desorb analyte into the instrument. Sampling, detection, identification, and quantitation of all compounds were readily achieved in times of less than 2 min, with detection limits ranging from 800 parts per trillion to 3 parts per million depending on the analyte. For all but one analyte, detection limits were well below (3.5-130 times below) permissible exposure limits. A linear dynamic range of 1-2 orders of magnitude was obtained over the concentration ranges studied (sub-ppbv to ppmv) for all analytes.
一种基于直线离子阱质量分析仪的微型手持式质谱仪已应用于空气中痕量有毒化合物的监测。该仪器由电池供电,便于携带且坚固耐用。我们预计它将用于公共安全、工业卫生和环境监测领域。对九种有毒工业化合物的气态样品进行了测试,这些化合物分别是光气、环氧乙烷、二氧化硫、丙烯腈、氯化氰、氰化氢、丙烯醛、甲醛和对硫磷。构建了一个吸附阱进样口,作为大气与质谱仪真空腔之间的接口。在吸附剂床上选择性收集分析物后,将吸附剂管抽真空,然后加热以将分析物解吸到仪器中。所有化合物的采样、检测、鉴定和定量在不到2分钟的时间内即可轻松完成,检测限根据分析物的不同在万亿分之800至百万分之3之间。除一种分析物外,所有分析物的检测限均远低于(低3.5至130倍)允许接触限值。在所研究的浓度范围内(亚皮克每立方米至百万分比体积),所有分析物的线性动态范围为1至2个数量级。