Suppr超能文献

一种可生物降解的自增强、自膨胀的X射线显影性聚-L,D-丙交酯螺旋形部分输尿管支架的引流及抗反流特性:一项实验研究

Drainage and antireflux characteristics of a biodegradable self-reinforced, self-expanding X-ray-positive poly-L,D-lactide spiral partial ureteral stent: an experimental study.

作者信息

Lumiaho Juha, Heino Antero, Kauppinen Tomi, Talja Martti, Alhava Esko, Välimaa Tero, Törmälä Pertti

机构信息

Department of Surgery, Kuopio University Hospital, Kuopio, Finland.

出版信息

J Endourol. 2007 Dec;21(12):1559-64. doi: 10.1089/end.2005.0085.

Abstract

PURPOSE

To evaluate the drainage and antireflux characteristics of a new self-expandable self-reinforced poly-L,D-lactide partial ureteral stent (SR-PLA 96) in an experimental model.

MATERIALS AND METHODS

Twelve dogs were used as experimental animals. A low-midline laparotomy and cystotomy were performed on all animals. In group A (six animals), 50-mm long SR-PLA 96 ureteral stents with a double-helical spiral design were inserted into both ureters, leaving the lower ends 2 cm above the ureterovesical junction. In group B (six animals), both ureters were stented with traditional pigtail stents (C-Flex) Double-J; Cook Urological), which were removed 8 weeks after surgery. Renal function and ureteral patency were evaluated by dynamic kidney imaging and urography examinations at 6 and 12 weeks postoperatively. The degrees of vesicoureteral reflux at two levels of the ureters and at the level of the renal pelvis were evaluated by nuclear voiding cystograms at 6 weeks.

RESULTS

The partial SR-PLA 96 stent design showed more favorable antireflux properties that the Double-J stent design. The degree of vesicoureteral reflux, reflected in an increase of nuclear enhancement at 6 weeks, was lower in the distal (7.9% +/- 14.7% v 63.2% +/- 17.3%; P < 0.05) and middle (6.1% +/- 8.1% v. 45.5% +/- 19.5%; P = 0.15) levels of the ureters as well as at the level of the renal pelvis (-3.4% +/- 3.6% v 6.2% +/- 3.9%; P = 0.65) than in the Double-J-stented ureters. No significant differences in renal function or ureteral patency were observed at 12 weeks after the Double-J stents had been removed and the SR-PLA 96 stents had fragmented.

CONCLUSION

A self-expandable, self-reinforced SR-PLA 96 partial ureteral stent showed more favorable antireflux properties than a Double-J stent.

摘要

目的

在实验模型中评估一种新型自膨胀自增强聚-L,D-丙交酯部分输尿管支架(SR-PLA 96)的引流和抗反流特性。

材料与方法

12只犬用作实验动物。对所有动物进行下腹正中剖腹术和膀胱切开术。在A组(6只动物)中,将具有双螺旋设计的50毫米长SR-PLA 96输尿管支架插入双侧输尿管,下端位于输尿管膀胱连接处上方2厘米处。在B组(6只动物)中,双侧输尿管置入传统猪尾支架(C-Flex双J;库克泌尿外科公司),术后8周取出。术后6周和12周通过动态肾脏成像和尿路造影检查评估肾功能和输尿管通畅情况。术后6周通过核素排尿膀胱造影评估输尿管两个水平及肾盂水平的膀胱输尿管反流程度。

结果

部分SR-PLA 96支架设计显示出比双J支架设计更有利的抗反流特性。6周时核增强增加所反映的膀胱输尿管反流程度,在输尿管远端(7.9%±14.7%对63.2%±17.3%;P<0.05)、中段(6.1%±8.1%对45.5%±19.5%;P = 0.15)以及肾盂水平(-3.4%±3.6%对6.2%±3.9%;P = 0.65)均低于双J支架置入的输尿管。在双J支架取出且SR-PLA 96支架已碎裂12周后,未观察到肾功能或输尿管通畅情况有显著差异。

结论

自膨胀自增强的SR-PLA 96部分输尿管支架显示出比双J支架更有利的抗反流特性。