一种短的可生物降解螺旋输尿管支架比双J支架具有更好的抗反流和引流性能。

A short biodegradable helical spiral ureteric stent provides better antireflux and drainage properties than a double-J stent.

作者信息

Lumiaho Juha, Heino Antero, Aaltomaa Sirpa, Välimaa Tero, Talja Martti

机构信息

Department of Surgery, Kuopio University Hospital, Kuopio, Finland.

出版信息

Scand J Urol Nephrol. 2011 Mar;45(2):129-33. doi: 10.3109/00365599.2010.544673. Epub 2011 Jan 12.

Abstract

OBJECTIVE

The use of ureteric stents is a standard treatment for the relief of ureter blockages for benign or malignant reasons. The most common stent design in clinical use is a double-J stent with coiled ends to avoid stent displacement. However, there are a number of complications associated with stent use. A double-J stent design bypasses the ureterovesical junction, enables bladder pressure reflection to the renal pelvis and causes vesicoureteral reflux (VUR). This may result in scarring and renal failure.

MATERIAL AND METHODS

An animal model was used to investigate whether VUR can be avoided in stented ureters using a short biodegradable partial helical spiral stent design that leaves the ureterovesical junction intact.

MATERIALS AND METHODS

Eight female pigs were used. Ureters on the left side were stented using a short helical spiral SR-PLGA stent (group A) and ureters on the right side using double-J stents (group B). Simulated voiding cystoureterography and standard intravenous urography examinations were performed on all eight animals at 4 weeks and on the remaining four animals at 8 weeks.

RESULTS

An SR-PLGA single coiled partial stent demonstrated superior drainage properties to a double-J stent at 4 weeks (p = 0.020). A marked but not statistically significant difference in favour of a SR-PLGA stent was also observed at 8 weeks (p = 0.102). A statistically significant difference was observed in VCUG findings in favour of group A at immediate postoperative control as well as in the 4 and 8 week follow-up studies (p = 0.011, p = 0.010, p = 0.046, respectively).

CONCLUSION

A self-expandable, SR-PLGA partial ureteric stent presented with superior drainage and antireflux properties compared to a double-J stent. The reflux commonly related to double-J stent use can be minimized by using a partial ureteric stent design.

摘要

目的

使用输尿管支架是缓解因良性或恶性原因导致的输尿管梗阻的标准治疗方法。临床使用中最常见的支架设计是带有盘绕末端的双J支架,以避免支架移位。然而,支架使用存在一些并发症。双J支架设计绕过输尿管膀胱连接部,使膀胱压力反射至肾盂并导致膀胱输尿管反流(VUR)。这可能导致瘢痕形成和肾衰竭。

材料与方法

使用动物模型研究采用短的可生物降解部分螺旋形支架设计使输尿管膀胱连接部保持完整时,能否避免支架置入输尿管后的VUR。

材料与方法

使用8只雌性猪。左侧输尿管使用短螺旋形SR-PLGA支架置入(A组),右侧输尿管使用双J支架置入(B组)。对所有8只动物在4周时进行模拟排尿膀胱输尿管造影和标准静脉肾盂造影检查,对其余4只动物在8周时进行检查。

结果

在4周时,SR-PLGA单盘绕部分支架显示出比双J支架更好的引流性能(p = 0.020)。在8周时也观察到有利于SR-PLGA支架的显著但无统计学意义的差异(p = 0.102)。在术后即刻对照以及4周和8周随访研究中,膀胱输尿管造影结果显示有利于A组的统计学显著差异(分别为p = 0.011、p = 0.010、p = 0.046)。

结论

与双J支架相比,可自膨胀的SR-PLGA部分输尿管支架具有更好的引流和抗反流性能。通过使用部分输尿管支架设计,可将与双J支架使用相关的反流降至最低。