Suppr超能文献

德克斯辅助右心室辅助装置流入插管的研发。

Development of the DexAide right ventricular assist device inflow cannula.

作者信息

Ootaki Yoshio, Saeed Diyar, Ootaki Chiyo, Akiyama Masatoshi, Fumoto Hideyuki, Horai Tetsuya, Dessoffy Raymond, Catanese Jacquelyn, Massiello Alex, Horvath David, Chen Ji-Feng, Benefit Stephen, Golding Leonard A R, Fukamachi Kiyotaka

机构信息

Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio, USA.

出版信息

ASAIO J. 2008 Jan-Feb;54(1):31-6. doi: 10.1097/MAT.0b013e31815b3ea4.

Abstract

Cannula design and cannulation site can pose major limitations to chronic pump implantations in animal studies. The aim of this study was to evaluate the biocompatibility of various inflow cannula designs for the DexAide right ventricular assist device (RVAD). The DexAide RVAD was implanted for intended durations of 14, 30, or 90 days in 19 animals (mean 20 +/- 11 days). Seven inflow cannula designs were evaluated: angled titanium conduit with caged tip (two cases); flexible polyurethane coated polyvinyl chloride (PVC) tube (one case); open ended titanium (one case); a titanium cannula with a flange (six cases); a cannula with a gelatin coated flange (five cases); a cannula with an angled flange (one case); and open ended titanium with two side holes (three cases). The open ended titanium inflow cannula with two side holes positioned through the diaphragmatic surface of the right ventricle (RV) via a right thoracotomy showed good biocompatibility for the chronic animal study. Other cannulae inserted into the infundibular portion of the RV via a left thoracotomy showed significant depositions. Gelatin coated inflow cannula had the advantage to prevent tissue growth around the inflow cannula. The DexAide RVAD pump itself showed good biocompatibility, although nonadherent depositions originating from the inflow cannulae were captured onto the primary impeller blades.

摘要

在动物研究中,插管设计和插管部位可能对慢性泵植入造成重大限制。本研究的目的是评估用于DexAide右心室辅助装置(RVAD)的各种流入插管设计的生物相容性。DexAide RVAD在19只动物中植入了预定的14、30或90天(平均20±11天)。评估了七种流入插管设计:带笼状尖端的成角度钛导管(2例);柔性聚氨酯涂层聚氯乙烯(PVC)管(1例);开口端钛管(1例);带法兰的钛插管(6例);带明胶涂层法兰的插管(5例);带成角度法兰的插管(1例);以及带两个侧孔的开口端钛管(3例)。通过右胸切开术经右心室(RV)膈肌表面定位的带两个侧孔的开口端钛流入插管在慢性动物研究中显示出良好的生物相容性。通过左胸切开术插入RV漏斗部的其他插管显示有明显的沉积物。明胶涂层流入插管具有防止组织在流入插管周围生长的优点。DexAide RVAD泵本身显示出良好的生物相容性,尽管源自流入插管的非粘附性沉积物被捕获在主叶轮叶片上。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验