Blood pressure changes associated with tilting in normotensive subjects: differences in response pattern as measured by oscillometry and auscultation.

作者信息

Sidery M B, Macdonald I A

机构信息

Department of Physiology and Pharmacology, University of Nottingham Medical School, UK.

出版信息

Clin Auton Res. 1991 Jun;1(2):161-6. doi: 10.1007/BF01826214.

Abstract

Three non-invasive instruments were used to measure blood pressure in the supine position and on tilting--a conventional and a random-zero sphygmomanometer, and an oscillometric device (Accutorr 1A). Twenty normotensives volunteered for the study. There was no statistically significant difference in systolic blood pressure and diastolic blood pressure measured by the conventional and random-zero sphygmomanometers in the supine position. There was a difference between these recordings and those of the Accutorr, with the Accutorr giving higher readings of systolic blood pressure (p less than 0.001), analysis of variance, 95% confidence interval of the difference between the Accutorr and the random-zero was 5.1-15.7 mmHg) and lower readings of diastolic blood pressure (p less than 0.0001, analysis of variance, 95% confidence interval of the difference between the Accutorr and the random-zero was -12.2- -2.2 mmHg). On tilting, the Accutorr showed an increase in systolic blood pressure while the other two machines did not (p less than 0.01, analysis of variance). By contrast, the Accutorr detected a smaller rise in diastolic blood pressure than with the other two instruments (p less than 0.05, analysis of variance). The difference between blood pressure measurements made in the supine position by the two different techniques, auscultation and oscillometry, might be expected. However, the two different techniques do not detect the same blood pressure responses to a change in posture.

摘要

使用三种非侵入性仪器在仰卧位和倾斜位测量血压,即传统血压计、随机零点血压计和示波测量装置(Accutorr 1A)。20名血压正常者自愿参与该研究。传统血压计和随机零点血压计在仰卧位测量的收缩压和舒张压无统计学显著差异。这些记录与Accutorr的记录存在差异,Accutorr测得的收缩压读数更高(p<0.001,方差分析,Accutorr与随机零点血压计之间差异的95%置信区间为5.1 - 15.7 mmHg),舒张压读数更低(p<0.0001,方差分析,Accutorr与随机零点血压计之间差异的95%置信区间为-12.2 - -2.2 mmHg)。在倾斜时,Accutorr显示收缩压升高,而其他两台仪器未出现这种情况(p<0.01,方差分析)。相比之下,Accutorr检测到的舒张压升高幅度小于其他两台仪器(p<0.05,方差分析)。通过听诊和示波测量这两种不同技术在仰卧位测量血压时存在差异是可以预料的。然而,这两种不同技术并未检测到相同的体位变化时的血压反应。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验