Muranaka Hiroyuki, Horiguchi Takayoshi, Usui Shuji, Ueda Yoshitake, Nakamura Osamu, Ikeda Fumiaki
Medical Engineering and Technology, Integrated Human Sciences Studies, Hiroshima International University, Higashi-Hiroshima, Hiroshima, Japan.
Magn Reson Med Sci. 2007;6(4):199-209. doi: 10.2463/mrms.6.199.
We evaluated radiofrequency (RF) heating of a humerus implant embedded in a gel phantom during magnetic resonance (MR) imaging for the specific absorption rate (SAR), angle between the implant and static magnetic field (B(0)), and position of the implant in the irradiation coil.
We embedded a stainless steel humerus implant 2 cm deep in tissue-equivalent loop and mass phantoms, placed it parallel to the static magnetic field of a 1.5T MR scanner, and recorded the temperatures of the implant surface with RF-transparent fiberoptic sensors. We measured rises in temperature at the tips of the implant by varying the SAR from 0.2 to 4.0 W/kg and evaluated RF heating of the implant for its angle to B(0) and its displacement along B(0) from the center of the RF irradiation coil.
RF heating was similar for the loop and mass phantoms because the eddy current flows through the periphery of both. As the SAR increased, the temperature at the implant tip increased, and there was a linear relationship between the SAR and temperature rise. The values were 6.4 degrees C at 2.0 W/kg and 12.7 degrees C at 4.0 W/kg. Rise in temperature decreased steeply as the angle between the implant and B(0) surpassed 45 degrees . In addition, as the implant was displaced from the center of the RF coil to both ends, the rise in temperature decreased.
The rise in temperature in deep tissue was estimated to be higher than 1.0 degrees C for SAR above 0.4 W/kg. RF heating was greatest when the implant was set parallel to B(0). In MR imaging of patients with implants, there is a risk of RF heating when the loop of the eddy current is formed inside the body.
我们在磁共振(MR)成像过程中,评估了嵌入凝胶体模中的肱骨植入物的射频(RF)加热情况,包括比吸收率(SAR)、植入物与静磁场(B(0))之间的夹角以及植入物在照射线圈中的位置。
我们将一个不锈钢肱骨植入物嵌入深度为2厘米的组织等效环形和质量体模中,使其与1.5T MR扫描仪的静磁场平行放置,并用射频透明光纤传感器记录植入物表面的温度。我们通过将SAR从0.2瓦/千克变化到4.0瓦/千克来测量植入物尖端的温度升高,并评估植入物在与B(0)的夹角以及沿B(0)方向从射频照射线圈中心的位移情况下的RF加热情况。
环形和质量体模的RF加热情况相似,因为涡流都流经两者的周边。随着SAR的增加,植入物尖端的温度升高,并且SAR与温度升高之间存在线性关系。在2.0瓦/千克时温度值为6.4摄氏度,在4.0瓦/千克时为12.7摄氏度。当植入物与B(0)之间的夹角超过45度时,温度升高急剧下降。此外,随着植入物从射频线圈中心向两端位移,温度升高也会下降。
对于SAR高于0.4瓦/千克的情况,深部组织中的温度升高估计高于1.0摄氏度。当植入物与B(0)平行设置时,RF加热最大。在植入物患者的MR成像中,当体内形成涡流环时存在RF加热的风险。