一种实时超声时域相关血流仪。I. 理论与设计。

A real-time ultrasound time-domain correlation blood flowmeter. I. Theory and design.

作者信息

Hein I A, Chen J T, Jenkins W K, O'Brien W R

机构信息

Illinois Univ., Urbana, IL.

出版信息

IEEE Trans Ultrason Ferroelectr Freq Control. 1993;40(6):768-75. doi: 10.1109/58.248222.

Abstract

A real-time ultrasound time-domain correlation (UTDC) blood flowmeter has been developed. Real-time performance has been achieved through the implementation of a custom-designed high-speed residue-number system (RNS) hardware correlator. The flowmeter is interfaced to a commercial ultrasound imager and can produce one-dimensional velocity versus range graphs at a rate of three per second. It has been validated in a blood flow phantom under a variety of conditions along with in vivo measurements in the human carotid artery. The theory of the time-domain correlation technique, design and implementation of flowmeter hardware, and the important correlation parameters which affect the performance of the flowmeter are described.

摘要

一种实时超声时域相关(UTDC)血流仪已被开发出来。通过实施定制设计的高速余数系统(RNS)硬件相关器实现了实时性能。该血流仪与商用超声成像仪相连,能够以每秒三张的速率生成一维速度与距离的图表。它已在各种条件下的血流模型中以及人体颈动脉的体内测量中得到验证。描述了时域相关技术的理论、血流仪硬件的设计与实现以及影响血流仪性能的重要相关参数。