Hein I A, O'Brien W R
Dept. of Electr. and Comput. Eng., Illinois Univ., Urbana, IL.
IEEE Trans Ultrason Ferroelectr Freq Control. 1993;40(6):776-85. doi: 10.1109/58.248223.
For Part I, see ibid., vol.40, no.6, p.768-775 (1993). The performance of a real-time ultrasound time-domain correlation (UTDC) flowmeter employing a high-speed residue number system correlator under various flow conditions for different correlation parameter settings is evaluated. Previous work has incorporated a weighted-averaging scheme to estimate the flow velocity over a wide number of echo signal spacings. The present evaluation indicates that a linear averaging of 50 adjacent echo signal pairs produces equivalent accuracy and precision at real-time processing speed as compared to previous non-real time systems incorporating weighted averaging.