Measurement of diagnostic electronic linear arrays by miniature hydrophone scanning.

作者信息

Ide M, Ohdaira E

机构信息

Dept. of Electron. and Commun. Eng., Musashi Inst. of Technol., Tokyo.

出版信息

IEEE Trans Ultrason Ferroelectr Freq Control. 1988;35(2):214-9. doi: 10.1109/58.4171.

Abstract

Ultrasonic pressures, intensities and power were computed using a calibrated miniature hydrophone scanning system. To accomplish these measurements, the electronically scanned beam was stopped, and raster scans were made with the hydrophone over a number of planes perpendicular to the beam axis. Ultrasonic power also was measured using a radiation force balance, and the results compared well with those obtained using the hydrophone. Spatial average-temporal average intensities for the stopped beam ranged from 0.13 to 3.71 mW/cm(2 ) (for a -40-dB beam area), and spatial peak-temporal peak intensities varied from 8.31 to 45.2 W/cm(2). Such measurements are important in setting equipment safety standards, in encouraging minimum use of ultrasonic exposures, and in quantifying exposure levels that may result in harmful biological effects.