Duck F A, Starritt H C, Aindow J D, Perkins M A, Hawkins A J
Br J Radiol. 1985 Oct;58(694):989-1001. doi: 10.1259/0007-1285-58-694-989.
A survey of the powers, pressures and intensities generated by ultrasonic pulse-echo equipment in clinical use has been carried out. Three conventional B-scanners, four linear-array scanners and four mechanically sectored scanners were included in the study. Measurements were made on a total of 22 transducers covering the nominal frequency range 2.25-7.5 MHz. On those instruments where an output power control was provided, two measurements were made: one at the maximum available power and a second at a lower power. On arrays with a variable transmit focus control, measurements were made at all available focus settings. In all, measurements were made on 38 separate focused pulsed ultrasonic fields. The measurements were carried out using a calibrated ultrasonic force balance, and a calibrated polyvinylidene difluoride (PVdF) membrane hydrophone. A very wide range of maximum powers, pressures and intensities were found. Powers from 0.5-80 mW were measured; spatial-average temporal-peak positive pressures at the transducer varied between 30 kPa and 1.15 MPa, and spatial-peak pulse-average intensities were in the range 3.6 X 10(3)-1.1 X 10(7) Wm-2.
已对临床使用的超声脉冲回波设备产生的功率、压力和强度进行了一项调查。该研究纳入了三台传统B型扫描仪、四台线性阵列扫描仪和四台机械扇形扫描仪。总共对22个换能器进行了测量,其标称频率范围为2.25 - 7.5MHz。对于那些设有输出功率控制的仪器,进行了两次测量:一次在最大可用功率下,另一次在较低功率下。对于具有可变发射聚焦控制的阵列,在所有可用的聚焦设置下进行了测量。总共对38个单独的聚焦脉冲超声场进行了测量。测量使用校准的超声力平衡仪和校准的聚偏二氟乙烯(PVdF)膜水听器进行。发现了非常广泛的最大功率、压力和强度范围。测量的功率为0.5 - 80mW;换能器处的空间平均时间峰值正压力在30kPa至1.15MPa之间变化,空间峰值脉冲平均强度在3.6×10³ - 1.1×10⁷Wm⁻²范围内。