Corneal aberrations and objective visual quality after hyperopic laser in situ keratomileusis using the Esiris excimer laser.

作者信息

Alió Jorge L, Piñero David P, Espinosa María José Ayala, Corral María José García

机构信息

Vissum-Instituto Oftalmológico de Alicante, Universidad Miguel Hernández, Alicante, Spain.

出版信息

J Cataract Refract Surg. 2008 Mar;34(3):398-406. doi: 10.1016/j.jcrs.2007.09.045.

Abstract

PURPOSE

To evaluate the changes in visual performance and ocular optical quality after hyperopic laser in situ keratomileusis (LASIK) using the Esiris excimer laser (Schwind eye-tech-solutions).

SETTING

Vissum-Instituto Oftalmológico de Alicante, Alicante, Spain.

METHODS

Excimer laser surgery was performed with the Esiris laser in 51 eyes of 29 patients with hyperopia or hyperopic astigmatism. The follow-up was 6 months. Changes in visual acuity, refraction, and corneal and ocular wavefront aberrations (6.0 mm pupil) were recorded and analyzed.

RESULTS

The mean preoperative sphere was +4.45 diopters (D)+/-1.08 (SD) (range +2.50 to +7.25 D) and the mean preoperative cylinder, -0.55+/-0.36 D (range 0.00 to 1.00 D). Ten eyes (19.61%) had a LASIK enhancement during the follow-up for the correction of the residual refractive error and were excluded from the refractive analysis. Of the 41 remaining eyes, 95.12% had the same or improved distance best spectacle-corrected visual acuity (BSCVA) 6 months after surgery; 90.25% had no change or a gain of lines of near BSCVA. Moreover, 80.50% of eyes were within +/-0.50 D of emmetropia. Regarding corneal aberrations, statistically significant changes were observed in higher-order aberrations (HOAs), coma-like aberrations, and the primary spherical aberration coefficient Z(4,0), which changed from positive to negative values (all P<.01). Regarding total ocular aberrations, statistically significant changes were observed in total and higher-order root-mean-square values (P<.01).

CONCLUSIONS

Hyperopic LASIK using the Esiris excimer laser for the correction of 2nd-order aberrations was safe and effective. Ocular and corneal HOAs increased significantly postoperatively.

摘要

目的

使用艾西丽准分子激光(施温德眼科技术解决方案公司)评估远视性准分子原位角膜磨镶术(LASIK)后视觉性能和眼屈光质量的变化。

设置

西班牙阿利坎特市阿利坎特眼科研究所Vissum。

方法

使用艾西丽激光对29例远视或远视散光患者的51只眼进行准分子激光手术。随访6个月。记录并分析视力、屈光、角膜和眼波前像差(6.0毫米瞳孔)的变化。

结果

术前平均球镜度数为+4.45屈光度(D)±1.08(标准差)(范围为+2.50至+7.25 D),术前平均柱镜度数为-0.55±0.36 D(范围为0.00至1.00 D)。10只眼(19.61%)在随访期间因矫正残余屈光不正而进行了LASIK增效手术,并被排除在屈光分析之外。在其余41只眼中,95.12%的患者术后6个月的远距离最佳矫正视力(BSCVA)相同或提高;90.25%患者的近距离BSCVA无变化或提高了行数。此外,80.50%的眼屈光状态在正视眼的±0.50 D范围内。关于角膜像差,高阶像差(HOAs)、类彗差和初级球差系数Z(4,0)有统计学意义的变化(均从正值变为负值,P<0.01)。关于总眼像差,总均方根值和高阶均方根值有统计学意义的变化(P<0.01)。

结论

使用艾西丽准分子激光进行远视性LASIK以矫正二阶像差是安全有效的。术后眼和角膜HOAs显著增加。