Pattanapanyasat Kovit, Sukapirom Kasama, Kowawisatsut Laddawan, Thepthai Charin
Center of Excellence for Flow Cytometry, Office for Research and Development, Faculty of Medicine, Siriraj Hospital, Mahidol University, Bangkok, Thailand.
Cytometry B Clin Cytom. 2008;74 Suppl 1:S98-106. doi: 10.1002/cyto.b.20415.
Absolute CD4+ T-lymphocyte counts are used in the initiation and monitoring of antiretroviral therapy in HIV-infected patients. Becton Dickinson's (BD) FACSCount system was introduced 12 years ago as a dedicated instrument for enumeration of absolute CD4+ T-lymphocytes. However, this system does not provide percent CD4+ T-lymphocyte that is the required monitoring parameter in pediatric patients. We evaluated a new BD FACSCount CD4 software and reagents for simultaneous percent and absolute CD4+ T-lymphocytes in HIV-infected blood.
Percent and absolute CD4+ T-lymphocytes in 149 HIV-infected blood samples were determined using a new FACSCount system. Results of percent and absolute CD4+ T-lymphocytes were compared between the dual-platform (DP) method, using BD FACScan flow cytometer plus hematology analyzer and the standard FACSCount system. Correlation and agreement were analyzed using linear regression and Bland-Altman analysis.
Percent CD4+ T-lymphocyte values obtained from the new FACSCount system correlated well with DP FACScan method (r2 = 0.977, P < 0.0001). Mean bias was only -0.36% [limit of agreement (LOA): -2.52% to +1.80%] and percent similarity was 101.36%. For absolute CD4+ T-lymphocyte, the new system correlated highly with standard FACSCount system (r2 = 0.986, P < 0.0001), with a percent similarity of 98.2. Mean bias was +3.39 cells/microl with LOA of -52.53 cells/mul to +59.31 cells/microl.
This new FACSCount system is a simple and reliable system for enumeration of absolute and percent CD4+ T-lymphocytes. Having one system giving both results should reduce the cost and thus increase access to CD4 testing for pediatric and adult patients.
绝对CD4 + T淋巴细胞计数用于启动和监测HIV感染患者的抗逆转录病毒治疗。12年前推出的贝克顿·迪金森公司(BD)的FACSCount系统是一种专门用于计数绝对CD4 + T淋巴细胞的仪器。然而,该系统无法提供儿科患者所需的监测参数——CD4 + T淋巴细胞百分比。我们评估了一种新的BD FACSCount CD4软件和试剂,用于同时检测HIV感染血液中的CD4 + T淋巴细胞百分比和绝对计数。
使用新的FACSCount系统测定149份HIV感染血液样本中的CD4 + T淋巴细胞百分比和绝对计数。将双平台(DP)方法(使用BD FACScan流式细胞仪和血液分析仪)与标准FACSCount系统所得的CD4 + T淋巴细胞百分比和绝对计数结果进行比较。采用线性回归和布兰德-奥特曼分析对相关性和一致性进行分析。
新FACSCount系统测得的CD4 + T淋巴细胞百分比值与DP FACScan方法相关性良好(r2 = 0.977,P < 0.0001)。平均偏差仅为-0.36% [一致性界限(LOA):-2.52%至+1.80%],相似百分比为101.36%。对于绝对CD4 + T淋巴细胞,新系统与标准FACSCount系统高度相关(r2 = 0.986,P < 0.0001),相似百分比为98.2。平均偏差为+3.39个细胞/微升,LOA为-52.53个细胞/微升至+59.31个细胞/微升。
这种新的FACSCount系统是一种简单可靠的系统,可用于计数绝对CD4 + T淋巴细胞和CD4 + T淋巴细胞百分比。使用一个系统就能得出两种结果,应可降低成本,从而增加儿科和成年患者进行CD4检测的机会。