用于同时计数HIV-1感染儿科患者中CD4 T淋巴细胞百分比和绝对数量的新型BD FACSCount CD4试剂系统。

New BD FACSCount CD4 reagent system for simultaneous enumeration of percent and absolute CD4 T-lymphocytes in HIV-1-infected pediatric patients.

作者信息

Pattanapanyasat Kovit, Sukapirom Kasama, Kowawisatsut Laddawan, Thepthai Charin

机构信息

Center of Excellence for Flow Cytometry, Office for Research and Development, Faculty of Medicine, Siriraj Hospital, Mahidol University, Bangkok, Thailand.

出版信息

Cytometry B Clin Cytom. 2008;74 Suppl 1:S98-106. doi: 10.1002/cyto.b.20415.

Abstract

BACKGROUND

Absolute CD4+ T-lymphocyte counts are used in the initiation and monitoring of antiretroviral therapy in HIV-infected patients. Becton Dickinson's (BD) FACSCount system was introduced 12 years ago as a dedicated instrument for enumeration of absolute CD4+ T-lymphocytes. However, this system does not provide percent CD4+ T-lymphocyte that is the required monitoring parameter in pediatric patients. We evaluated a new BD FACSCount CD4 software and reagents for simultaneous percent and absolute CD4+ T-lymphocytes in HIV-infected blood.

METHODS

Percent and absolute CD4+ T-lymphocytes in 149 HIV-infected blood samples were determined using a new FACSCount system. Results of percent and absolute CD4+ T-lymphocytes were compared between the dual-platform (DP) method, using BD FACScan flow cytometer plus hematology analyzer and the standard FACSCount system. Correlation and agreement were analyzed using linear regression and Bland-Altman analysis.

RESULTS

Percent CD4+ T-lymphocyte values obtained from the new FACSCount system correlated well with DP FACScan method (r2 = 0.977, P < 0.0001). Mean bias was only -0.36% [limit of agreement (LOA): -2.52% to +1.80%] and percent similarity was 101.36%. For absolute CD4+ T-lymphocyte, the new system correlated highly with standard FACSCount system (r2 = 0.986, P < 0.0001), with a percent similarity of 98.2. Mean bias was +3.39 cells/microl with LOA of -52.53 cells/mul to +59.31 cells/microl.

CONCLUSION

This new FACSCount system is a simple and reliable system for enumeration of absolute and percent CD4+ T-lymphocytes. Having one system giving both results should reduce the cost and thus increase access to CD4 testing for pediatric and adult patients.

摘要

背景

绝对CD4 + T淋巴细胞计数用于启动和监测HIV感染患者的抗逆转录病毒治疗。12年前推出的贝克顿·迪金森公司(BD)的FACSCount系统是一种专门用于计数绝对CD4 + T淋巴细胞的仪器。然而,该系统无法提供儿科患者所需的监测参数——CD4 + T淋巴细胞百分比。我们评估了一种新的BD FACSCount CD4软件和试剂,用于同时检测HIV感染血液中的CD4 + T淋巴细胞百分比和绝对计数。

方法

使用新的FACSCount系统测定149份HIV感染血液样本中的CD4 + T淋巴细胞百分比和绝对计数。将双平台(DP)方法(使用BD FACScan流式细胞仪和血液分析仪)与标准FACSCount系统所得的CD4 + T淋巴细胞百分比和绝对计数结果进行比较。采用线性回归和布兰德-奥特曼分析对相关性和一致性进行分析。

结果

新FACSCount系统测得的CD4 + T淋巴细胞百分比值与DP FACScan方法相关性良好(r2 = 0.977,P < 0.0001)。平均偏差仅为-0.36% [一致性界限(LOA):-2.52%至+1.80%],相似百分比为101.36%。对于绝对CD4 + T淋巴细胞,新系统与标准FACSCount系统高度相关(r2 = 0.986,P < 0.0001),相似百分比为98.2。平均偏差为+3.39个细胞/微升,LOA为-52.53个细胞/微升至+59.31个细胞/微升。

结论

这种新的FACSCount系统是一种简单可靠的系统,可用于计数绝对CD4 + T淋巴细胞和CD4 + T淋巴细胞百分比。使用一个系统就能得出两种结果,应可降低成本,从而增加儿科和成年患者进行CD4检测的机会。