Immunoassay interference causing unexpected reactive results in magnetic-microparticle-based assays was detected. A systematic evaluation of Liaison Epstein-Barr virus immunoglobulin M showed that 5% of the positive results (0.4% of tested samples) could be explained by such interference. Adding chemical blocking reagents (polyvinylpyrrolidone and polyvinyl alcohol) to the assay buffers partially prevented this phenomenon.
检测到免疫测定干扰在基于磁性微粒的测定中导致意外的反应性结果。对Liaison Epstein-Barr病毒免疫球蛋白M的系统评估表明,5%的阳性结果(0.4%的测试样本)可归因于这种干扰。在测定缓冲液中添加化学封闭试剂(聚乙烯吡咯烷酮和聚乙烯醇)可部分防止这种现象。