Delaney Jeffrey W, Li Jennifer S, Rhodes John F
Duke University Medical Center, Division of Pediatric Cardiology, Durham, NC 27710, USA.
Congenit Heart Dis. 2007 Jul-Aug;2(4):256-64. doi: 10.1111/j.1747-0803.2007.00107.x.
To summarize major complications and outcome for patients receiving percutaneous closure of atrial septal communications.
The Medline database and the United States Food and Drug Administration manufacturer and user facility device experience databases (MAUDE) were searched for reports related to complications with atrial septal occluding devices. The medical literature documenting complication rates for these devices were reviewed and summarized. The MAUDE database complication reports were compared with those reported in the medical literature using national implant estimates.
The MAUDE database correlated in the type of complications most frequently encountered with each device. However, based on estimated total implant numbers, there is a higher incidence of major complications, including death. AGA devices had a 0.3% erosion/perforation rate with a higher morbidity and mortality (29%) than previously reported. NMT devices had a lower incidence of erosion/perforation rate of 0.05%. Embolization rates for the NMT devices were also lower than published European studies, possibly reflecting the US restriction of the device for closure of patent foramen ovale. Thrombus was more frequently encountered on the NMT device. Both AGA and NMT devices have been shown to be safe and effective alternatives to cardiac surgery. The MAUDE database correlated, with a very low overall complication rate, but showed a higher estimated major complication rate than the medical literature. These data demonstrate the difficulty in quantifying rare complications in the premarketing analysis and the obligation providers have to report and evaluate complications through vigilant postmarketing surveillance.
总结接受经皮封堵房间隔缺损患者的主要并发症及预后情况。
检索Medline数据库以及美国食品药品监督管理局制造商和用户设施设备经验数据库(MAUDE),查找与房间隔封堵装置并发症相关的报告。对记录这些装置并发症发生率的医学文献进行回顾和总结。使用全国植入估计数,将MAUDE数据库中的并发症报告与医学文献中报告的进行比较。
MAUDE数据库与每种装置最常遇到的并发症类型相关。然而,根据估计的总植入数量,包括死亡在内的主要并发症发生率更高。AGA装置的侵蚀/穿孔率为0.3%,其发病率和死亡率(29%)高于先前报告。NMT装置的侵蚀/穿孔率较低,为0.05%。NMT装置的栓塞率也低于已发表的欧洲研究,这可能反映了美国对该装置用于封堵卵圆孔未闭的限制。NMT装置上血栓更常见。AGA和NMT装置均已被证明是心脏手术安全有效的替代方法。MAUDE数据库显示总体并发症发生率非常低,但估计的主要并发症发生率高于医学文献。这些数据表明在上市前分析中量化罕见并发症存在困难,以及供应商有义务通过警惕的上市后监测来报告和评估并发症。