Mahoney J F, Perel J, Ruatta S A, Martino P A, Husain S, Lee T D
Phrasor Scientific Inc., Duarte, CA 91010.
Rapid Commun Mass Spectrom. 1991 Oct;5(10):441-5. doi: 10.1002/rcm.1290051004.
A new ion desorption method is described that utilizes a primary beam of massive, multiply charged cluster ions to generate secondary ions of peptides in a glycerol matrix. The massive cluster ion beam is generated via electrohydrodynamic emission using a 1.5 M solution of ammonium acetate in 30% aqueous glycerol. Negative ion spectra of peptides obtained using this technique show greatly decreased relative intensities for fragment ions and 'chemical noise' background when compared to spectra obtained using a xenon atom primary beam. The near absence of fragments derived from radiation damage to the sample solution is attributed to the impact of primary particles with energies less than 1 eV/nucleon.