通过共振频率分析测量确定的牙种植体稳定性与种植体周围垂直骨缺损之间的关系:一项体外研究。

Relationship between dental implant stability determined by resonance frequency analysis measurements and peri-implant vertical defects: an in vitro study.

作者信息

Tözüm T F, Turkyilmaz I, McGlumphy E A

机构信息

Department of Periodontology, Faculty of Dentistry, Hacettepe University, Ankara, Turkey.

出版信息

J Oral Rehabil. 2008 Oct;35(10):739-44. doi: 10.1111/j.1365-2842.2007.01840.x. Epub 2008 Apr 14.

Abstract

Resonance frequency (RF) analysis is a non-invasive and objective technique developed for implantology, where it measures the stability of the implant in the bone socket. A limited number of studies were performed to compare different stability measurement devices. The aim of the study was to evaluate the association between the implant stability and peri-implant vertical defect by using the new wireless RF analyser. Micro-textured rough-surfaced implants with diameters of 3.7 and 4.8 mm and with a length of 12 mm were used. The peak insertion torque was recorded with the help of the torque control system only during the implant placement. The peri-implant defects were created in millimetre increments ranging between 0 and 5 mm to the same extent on all implants. The RF analysis values were measured by using the RF analyser. Significant correlation was found between insertion torque (44.3 +/- 3 N cm) and implant stability quotient (ISQ) (all implants: r = 0.76, 3.7 mm diameter: r = 0.65, 4.8 mm diameter: r = 0.80). For 3.7 x 12-mm(2) implants, the mean ISQ values were 72.6 +/- 2.4, 69.8 +/- 2, 67.2 +/- 1.8, 64.3 +/- 1.9, 61.2 +/- 2 and 57.2 +/- 2.5 when peri-implant vertical defects were 0, 1, 2, 3, 4 and 5 mm; and the corresponding values were 76.5 +/- 2.2, 74.5 +/- 2.2, 72.3 +/- 2, 70 +/- 1.9, 67 +/- 1.9 and 63.4 +/- 2.3, respectively, for 4.8 x 12-mm(2) implants. All values were significantly different when compared with each other. The wireless RF analyser seems to be a suitable and reliable device to determine the implant stability. Peri-implant bone loss simulated by using acrylic models may result with a decrease in ISQ values for osseo-integrated implants measured by the RF analyser.

摘要

共振频率(RF)分析是一种为种植牙学开发的非侵入性客观技术,用于测量种植体在牙槽骨中的稳定性。仅有少数研究对不同的稳定性测量设备进行了比较。本研究的目的是使用新型无线RF分析仪评估种植体稳定性与种植体周围垂直骨缺损之间的关联。使用了直径为3.7毫米和4.8毫米、长度为12毫米的微纹理粗糙表面种植体。仅在种植体植入过程中借助扭矩控制系统记录峰值植入扭矩。在所有种植体上以0至5毫米的毫米增量同等程度地制造种植体周围骨缺损。使用RF分析仪测量RF分析值。发现植入扭矩(44.3±3牛厘米)与种植体稳定性商数(ISQ)之间存在显著相关性(所有种植体:r = 0.76,直径3.7毫米:r = 0.65,直径4.8毫米:r = 0.80)。对于3.7×12毫米²的种植体,当种植体周围垂直骨缺损为0、1、2、3、4和5毫米时,平均ISQ值分别为72.6±2.4、69.8±2、67.2±1.8、64.3±1.9、61.2±2和57.2±2.5;对于4.8×12毫米²的种植体,相应的值分别为76.5±2.2、74.5±2.2、72.3±2、70±1.9、67±1.9和63.4±2.3。所有值相互比较时均有显著差异。无线RF分析仪似乎是确定种植体稳定性的合适且可靠的设备。使用丙烯酸模型模拟的种植体周围骨丢失可能导致通过RF分析仪测量的骨结合种植体的ISQ值降低。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验