新鲜拔牙窝内种植体的缺损深度对共振频率分析及植入扭矩值的影响:一项人体尸体研究

Influence of defect depth on resonance frequency analysis and insertion torque values for implants placed in fresh extraction sockets: a human cadaver study.

作者信息

Turkyilmaz Ilser, Sennerby Lars, Yilmaz Burak, Bilecenoglu Burak, Ozbek Esma Nida

机构信息

Department of Restorative and Prosthetic Dentistry, College of Dentistry, The Ohio State University, Columbus, OH 43218-2357, USA.

出版信息

Clin Implant Dent Relat Res. 2009 Mar;11(1):52-8. doi: 10.1111/j.1708-8208.2008.00095.x. Epub 2008 Apr 1.

Abstract

BACKGROUND

Clinical studies show promising outcomes with implants inserted at the time of extraction. However, this often results in an initial bone defect at the marginal region which preferably should heal for an optimal function. Therefore, monitoring of these implants is vital.

PURPOSES

The aims of this study were to determine the initial stability of implants placed into fresh extraction sockets, and to explore the correlations between the peri-implant bone levels and implant stability parameters.

MATERIALS AND METHODS

Six human cadaver mandibles including all natural teeth were selected for this study. All natural teeth were gently extracted, and 84 implants were immediately placed into fresh extraction sockets with five different implant depths. The maximum insertion torque values were recorded, and primary implant stability measurements were performed by means of resonance frequency analysis (RFA). The vertical distance between implant/abutment junction and the first bone-implant contact was recorded using a periodontal probe.

RESULTS

It was found that the insertion torque and RFA were 28.9 +/- 7 Ncm and 65.6 +/- 9 implant stability quotient (ISQ), respectively, for 420 measurements from all 84 implants. Statistically significant correlation was found between insertion torque and ISQ values (r = 0.86; p < .001) for all implants. Both insertion torque and ISQ values dramatically decreased when the amount of peri-implant vertical bone defect increased.

CONCLUSION

The results of this study demonstrated a linear relationship between peri-implant vertical bone defect depth and RFA value. It is proposed that the RFA method is sensitive to detect changes of the marginal bone level and may be used to monitor healing of peri-implant bone defects.

摘要

背景

临床研究表明,拔牙时同期植入种植体可取得良好效果。然而,这通常会在边缘区域形成初始骨缺损,而该区域最好愈合以实现最佳功能。因此,对这些种植体进行监测至关重要。

目的

本研究旨在确定植入新鲜拔牙窝的种植体的初始稳定性,并探讨种植体周围骨水平与种植体稳定性参数之间的相关性。

材料与方法

本研究选取了六具包含所有天然牙的人类尸体下颌骨。所有天然牙均被轻柔拔除,84枚种植体即刻被植入新鲜拔牙窝,植入深度分为五种不同情况。记录最大植入扭矩值,并通过共振频率分析(RFA)进行种植体初始稳定性测量。使用牙周探针记录种植体/基台连接处与首次骨-种植体接触之间的垂直距离。

结果

从所有84枚种植体的420次测量中发现,植入扭矩和RFA分别为28.9±7 Ncm和65.6±9种植体稳定性商数(ISQ)。所有种植体的植入扭矩和ISQ值之间存在统计学上的显著相关性(r = 0.86;p <.001)。当种植体周围垂直骨缺损量增加时,植入扭矩和ISQ值均显著降低。

结论

本研究结果表明种植体周围垂直骨缺损深度与RFA值之间存在线性关系。建议RFA方法对检测边缘骨水平变化敏感,可用于监测种植体周围骨缺损的愈合情况。