Freise K J, Schmidt R L, Gingerich E L, Veng-Pedersen P, Widness J A
College of Pharmacy, The University of Iowa, Iowa City, IA 52242, USA.
Int J Lab Hematol. 2009 Oct;31(5):496-504. doi: 10.1111/j.1751-553X.2008.01066.x. Epub 2008 Apr 14.
The objective of the study was to determine whether selected hematologic parameters measured on umbilical cord blood samples using an automated hematology analyzer (Sysmex XE-2100) were affected by (i) anticoagulant (the specimens were collected in EDTA vs. sodium heparin), (ii) temperature (the specimens were maintained at 4 degrees C vs. room temperature for up to 72 h) and (iii) 1 : 5 dilution vs. undiluted using the manufacturer's diluting solution. Use of heparin, instead of EDTA, had little effect on the hematologic results (n = 8) except for lower platelet and progenitor cell counts. Results were remarkably stable for 72 h at either room temperature or 4 degrees C except for modest red blood cell swelling at 24 h. Specimens of blood diluted at 1 : 5 had an immediate small, but significant change on white cell count (+13.3%), reticulocyte count (-11.2%) and reticulocyte hemoglobin content (-19.6%). Diluted samples did not change further over 4 h at room temperature. With a 1 : 5 dilution, analysis of 40 microl of cord blood stored for 3 days at room temperature may provide useful hematologic information with little phlebotomy loss.
本研究的目的是确定使用自动血液分析仪(Sysmex XE - 2100)检测脐带血样本时,选定的血液学参数是否受以下因素影响:(i)抗凝剂(样本采集于乙二胺四乙酸(EDTA)或肝素钠中);(ii)温度(样本在4℃或室温下保存长达72小时);以及(iii)使用制造商的稀释液进行1:5稀释与未稀释。除血小板和祖细胞计数较低外,使用肝素而非EDTA对血液学结果影响不大(n = 8)。除24小时时红细胞有适度肿胀外,样本在室温或4℃下72小时内结果均非常稳定。1:5稀释的血液样本白细胞计数立即出现微小但显著的变化(+13.3%)、网织红细胞计数(-11.2%)和网织红细胞血红蛋白含量(-19.6%)。稀释后的样本在室温下4小时内未进一步变化。采用1:5稀释时,对室温下保存3天的40微升脐带血进行分析,可能在几乎不损失静脉穿刺血量的情况下提供有用的血液学信息。