Hamann M, Aldridge M, Dickson J, Endozo R, Lozhkin K, Hutton B F
Institute of Nuclear Medicine-UCL, London, UK.
Phys Med Biol. 2008 May 21;53(10):2495-508. doi: 10.1088/0031-9155/53/10/003. Epub 2008 Apr 18.
The 4-slice CT that forms part of the GE Infinia Hawkeye-4 SPECT-CT scanner (Hawkeye) is evaluated against the diagnostic 16-slice CT that is incorporated in the GE Discovery ST PET-CT system (DST). The x-ray tube of the slow-rotating Hawkeye system (23 s/rotation) operates at approximately a third of the dose of diagnostic systems as used for conventional diagnostic imaging. Image reconstruction is optimized for low noise. High-contrast spatial resolution significantly falls behind diagnostic figures: the average of MTF(50) and MTF(10) (resolution where the MTF has fallen to 50% and 10%) is 2.8 +/- 0.1 cm(-1) for Hawkeye and 5.3 +/- 0.1 cm(-1) for the DST (standard reconstruction filters). Resolution in the direction of the couch movement (z coordinate) is governed by the fixed Hawkeye slice width of 5 mm. Reconstruction accuracy is found to be increased by reducing the default z increment from 4.4 mm to 2.2 mm. Low-contrast object detectability is superior compared with diagnostic systems operating in the Hawkeye dose range. In the diagnostic dose regime, however, small low-contrast details remain visible in DST that are not detectable with Hawkeye. Although not of diagnostic quality, the low-dose Hawkeye provides appropriate data for SPECT attenuation correction and anatomical localization capability.
对通用电气公司Infinia Hawkeye - 4单光子发射计算机断层扫描 - X线计算机断层扫描(SPECT - CT)扫描仪(Hawkeye)中包含的4层CT与通用电气公司Discovery ST正电子发射断层显像 - X线计算机断层扫描(PET - CT)系统(DST)中包含的诊断用16层CT进行了评估。慢速旋转的Hawkeye系统(23秒/旋转)的X线管运行时的剂量约为用于传统诊断成像的诊断系统剂量的三分之一。图像重建针对低噪声进行了优化。高对比度空间分辨率明显低于诊断数值:对于Hawkeye,调制传递函数(MTF)(50)和MTF(10)(MTF降至50%和10%时的分辨率)的平均值为2.8±0.1厘米⁻¹,对于DST为5.3±0.1厘米⁻¹(标准重建滤波器)。沿检查床移动方向(z坐标)的分辨率由Hawkeye固定的5毫米层厚决定。发现将默认的z增量从4.4毫米减少到2.2毫米可提高重建精度。与在Hawkeye剂量范围内运行的诊断系统相比,低对比度物体的可检测性更优。然而,在诊断剂量模式下,DST中仍可见一些小的低对比度细节,而Hawkeye无法检测到这些细节。尽管Hawkeye的图像质量未达到诊断标准,但低剂量的Hawkeye可为SPECT衰减校正和解剖定位能力提供合适的数据。