Weir Victor J, Zhang Jie, Bruner Angela P
Baylor Health Care System, Suite Dallas, TX, USA.
Department of Radiology, University of Kentucky, Lexington, KY, USA.
J Xray Sci Technol. 2015;23(3):373-81. doi: 10.3233/XST-150496.
Radiation dose and image quality from a recently introduced mobile CT imaging system are presented. Radiation dose was measured using a conventional 100 mm pencil ionization chamber and CT polymethylmetacrylate (PMMA) body and head phantoms. Image quality was evaluated with a CATPHAN 500 phantom. Spatial resolution, low contrast resolution, Modulation Transfer Function (MTF), and Normalized Noise Power Spectrum (NNPS) were analyzed. Radiation dose and image quality were compared to those from a multi-detector CT scanner (Siemens Sensation 64). Under identical technique factors radiation dose (mGy/mAs) from the AIRO mobile CT system (AIRO) is higher than that from a 64 slice CT scanner. Based on MTF analysis, both Soft and Standard filters of the AIRO system lost resolution quickly compared to the Sensation 64 slice CT. The Siemens scanner had up to 7 lp/cm for the head FOV and H40 kernel and up to 5 lp/cm at body FOV for the B40f kernel. The Standard kernel in the AIRO system was evaluated to have 3 lp/cm and 4 lp/cm for the body and head FOVs respectively. NNPS of the AIRO shows low frequency noise due to ring-like artifacts which may be caused by detector calibration or lack of artifact reducing image post-processing. Due to a higher dose in terms of mGy/mAs at both head and body FOV, the contrast to noise ratio is higher in the AIRO system than in the Siemens scanner. However detectability of the low contrast objects is poorer in the AIRO due to the presence of ring artifacts in the location of the targets.
介绍了一种最近推出的移动CT成像系统的辐射剂量和图像质量。使用传统的100毫米铅笔电离室以及CT聚甲基丙烯酸甲酯(PMMA)体模和头部体模测量辐射剂量。使用CATPHAN 500体模评估图像质量。分析了空间分辨率、低对比度分辨率、调制传递函数(MTF)和归一化噪声功率谱(NNPS)。将辐射剂量和图像质量与多探测器CT扫描仪(西门子Sensation 64)的进行了比较。在相同的技术参数下,AIRO移动CT系统(AIRO)的辐射剂量(mGy/mAs)高于64层CT扫描仪。基于MTF分析,与Sensation 64层CT相比,AIRO系统的软滤波器和标准滤波器的分辨率都迅速下降。对于头部视野和H40内核,西门子扫描仪的分辨率高达7 lp/cm,对于体部视野和B40f内核,分辨率高达5 lp/cm。AIRO系统中的标准内核在体部视野和头部视野中的分辨率分别评估为3 lp/cm和4 lp/cm。AIRO的NNPS显示出由于环状伪影导致的低频噪声,这可能是由探测器校准或缺乏减少伪影的图像后处理引起的。由于在头部和体部视野中mGy/mAs方面的剂量较高,AIRO系统中的对比度噪声比高于西门子扫描仪。然而,由于目标位置存在环状伪影,AIRO中低对比度物体的可检测性较差。