Mechalakos James G, Hunt Margie A, Lee Nancy Y, Hong Linda X, Ling C Clifton, Amols Howard I
Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, New York, U.S.A.
Department of Radiation Oncology, Memorial Sloan-Kettering Cancer Center, New York, New York, U.S.A.
J Appl Clin Med Phys. 2007 Sep 24;8(4):28-44. doi: 10.1120/jacmp.v8i4.2439.
The purpose of the present study was to use a kilovoltage imaging device to measure interfractional and intrafractional setup deviations in patients with head-and-neck or brain cancers receiving intensity-modulated radiotherapy (IMRT) treatment. Before and after IMRT treatment, approximately 3 times weekly, 7 patients were imaged using the Varian On-Board Imager (OBI: Varian Medical Systems, Palo Alto, CA), a kilovoltage imaging device permanently mounted on the gantry of a Varian 21EX LINAC (Varian Medical Systems). Because of commissioning of the remote couch correction of the OBI during the study, online setup corrections were performed on 2 patients. For the other 5 patients, weekly corrections were made based on a sliding average of the measured data. From these data, we determined the interfractional setup deviation (defined as the shift from the original setup position suggested by the daily image), the residual error associated with the weekly correction protocol, and the intrafractional setup deviation, defined as the difference between the post-treatment and pretreatment images. We also used our own image registration software to determine interfractional and intrafractional rotational deviations from the images based on the template-matching method. In addition, we evaluated the influence of inter-observer variation on our results, and whether the use of various registration techniques introduced differences. Finally, translational data were compared with rotational data to search for correlations. Translational setup errors from all data were 0.0 +/- 0.2 cm, -0.1 +/- 0.3 cm, and -0.2 +/- 0.3 cm in the right-left (RL), anterior-posterior (AP), and superior-inferior (SI) directions respectively. Residual error for the 5 patients with a weekly correction protocol was -0.1 +/- 0.2 cm (RL), 0.0 +/- 0.3 cm (AP), and 0.0 +/- 0.2 cm (SI). Intrafractional translation errors were small, amounting to 0.0 +/- 0.1 cm, -0.1 +/- 0.2 cm, and 0.0 +/- 0.1 cm in the RL, AP, and SI directions respectively. In the sagittal and coronal views respectively, interfractional rotational errors were -1.1 +/- 1.7 degrees and -0.5 +/- 0.9 degrees, and intrafractional rotational errors were 0.3 +/- 0.6 degrees and 0.2 +/- 0.5 degrees. No significant correlation was seen between translational and rotational data. The OBI image data were used to study setup error in the head-and-neck patients. Nonzero systematic errors were seen in the interfractional translational and rotational data, but not in the intrafractional data, indicating that the mask is better at maintaining head position than at reproducing it.
本研究的目的是使用千伏成像设备测量接受调强放射治疗(IMRT)的头颈癌或脑癌患者的分次间和分次内摆位偏差。在IMRT治疗前后,每周约3次,使用瓦里安机载成像仪(OBI:瓦里安医疗系统公司,加利福尼亚州帕洛阿尔托)对7例患者进行成像,该千伏成像设备永久安装在瓦里安21EX直线加速器(瓦里安医疗系统公司)的机架上。由于在研究期间对OBI的远程治疗床校正进行了调试,对2例患者进行了在线摆位校正。对于其他5例患者,根据测量数据的滑动平均值进行每周校正。根据这些数据,我们确定了分次间摆位偏差(定义为与每日图像建议的原始摆位位置的偏移)、与每周校正方案相关的残余误差以及分次内摆位偏差,定义为治疗后和治疗前图像之间的差异。我们还使用自己的图像配准软件,基于模板匹配方法从图像中确定分次间和分次内的旋转偏差。此外,我们评估了观察者间差异对结果的影响,以及使用各种配准技术是否会引入差异。最后,将平移数据与旋转数据进行比较以寻找相关性。所有数据在左右(RL)、前后(AP)和上下(SI)方向的平移摆位误差分别为0.0±0.2 cm、-0.1±0.3 cm和-0.2±0.3 cm。采用每周校正方案的5例患者的残余误差在RL方向为-0.1±0.2 cm,AP方向为0.0±0.3 cm,SI方向为0.0±0.2 cm。分次内平移误差较小,在RL、AP和SI方向分别为0.0±0.1 cm、-0.1±0.2 cm和0.0±0.1 cm。在矢状面和冠状面视图中,分次间旋转误差分别为-1.1±1.7度和-0.5±0.9度,分次内旋转误差分别为0.3±0.6度和0.2±0.5度。平移和旋转数据之间未发现显著相关性。OBI图像数据用于研究头颈患者的摆位误差。在分次间平移和旋转数据中观察到非零系统误差,但在分次内数据中未观察到,这表明面罩在保持头部位置方面比再现头部位置更好。