Department of Radiation Oncology, Shengjing Hospital of China Medial University, Shenyang 110022, China.
J Appl Clin Med Phys. 2012 Sep 6;13(5):3766. doi: 10.1120/jacmp.v13i5.3766.
The purpose of this study was to investigate the impact of different CBCT imaging monitor units (MUs), reconstruction slice thicknesses, and planning CT slice thicknesses on the positioning accuracy of a megavoltage cone-beam computed tomography (MV-CBCT) system in image-guided radiation therapy (IGRT) in head-and-neck patients. The MV-CBCT system was a Siemens MVision, a commercial system integrated into the Siemens ONCOR linear accelerator. The positioning accuracy of the MV-CBCT system was determined using an anthropomorphic phantom while varying the MV-CBCT imaging MU, reconstruction slice thickness, and planning CT slice thickness. A total of 240 CBCT images from six head-and-neck patients who underwent intensity-modulated radiotherapy (IMRT) treatment were acquired and reconstructed using different MV-CBCT scanning protocols. The interfractional setup errors of the patients were retrospectively analyzed for different imaging MUs, reconstruction slice thicknesses, and planning CT slice thicknesses. Using the anthropomorphic phantom, the largest measured mean deviation component and standard deviation of the MVision in 3D directions were 1.3 and 1.0 mm, respectively, for different CBCT imaging MUs, reconstruction slice thicknesses, and planning CT slice thicknesses. The largest setup group system error (M), system error (Σ), and random error (σ) from six head-and-neck patients were 0.6, 1.2, and 1.7 mm, respectively. No significant difference was found in the positioning accuracy of the MV-CBCT system between the 5 and 8 MUs, and between the 1 and 3 mm reconstruction slice thicknesses. A thin planning CT slice thickness may achieve higher positioning precision using the phantom measurement, but no significant difference was found in clinical setup precision between the 1 and 3 mm planning CT slice thicknesses.
本研究旨在探讨不同锥形束 CT(CBCT)成像监测单位(MU)、重建层厚和计划 CT 层厚对头部和颈部患者图像引导放射治疗(IGRT)中兆伏锥形束 CT(MV-CBCT)系统定位精度的影响。MV-CBCT 系统为西门子 MVision,是集成在西门子 ONCOR 直线加速器中的商用系统。通过改变 MV-CBCT 成像 MU、重建层厚和计划 CT 层厚,使用人体模型来确定 MV-CBCT 系统的定位精度。对 6 名接受调强放疗(IMRT)治疗的头颈部患者的 240 张 CBCT 图像进行了采集和重建,使用不同的 MV-CBCT 扫描方案。对头颈部患者不同成像 MU、重建层厚和计划 CT 层厚的分次间摆位误差进行了回顾性分析。在不同的 CBCT 成像 MU、重建层厚和计划 CT 层厚下,使用人体模型测量 MVision 的最大三维方向平均偏差分量和标准偏差分别为 1.3mm 和 1.0mm。6 名头颈部患者的最大总体系统误差(M)、系统误差(Σ)和随机误差(σ)分别为 0.6mm、1.2mm 和 1.7mm。在 5MU 和 8MU 之间,1mm 和 3mm 重建层厚之间,MV-CBCT 系统的定位精度没有显著差异。在体模测量中,薄的计划 CT 层厚可能达到更高的定位精度,但在临床摆位精度方面,1mm 和 3mm 计划 CT 层厚之间没有显著差异。