Suppr超能文献

[Influence of mechanical effect due to MRI-magnet on tattoo seal and eye makeup].

作者信息

Morishita Yuta, Miyati Tosiaki, Ueda Jousei, Shimizu Mitsuru, Hamaguchi Takashi, Fujiwara Yasuhiro, Hayashi Hiroyuki

机构信息

School of Health Sciences, Faculty of Medicine, Kanazawa University.

出版信息

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2008 May 20;64(5):587-90. doi: 10.6009/jjrt.64.587.

Abstract

The purpose of our study was to assess the mechanical effect on tattoo seals and eye makeup caused by a spatial magnetic gradient in the magnetic resonance imaging (MRI) system. Seven kinds of tattoo seals and three kinds of eye makeup, i.e., mascara, eye shadow, and eyeliner were used. On a 3.0-Tesla MRI, we determined these deflection angles according to a method established by the American Society for Testing and Materials (ASTM) at the position that produced the greatest magnetically induced deflection. Eighty-five percent of the tattoo seals showed deflection angles greater than 45 degrees of the ASTM guidelines, and the mascara and eye shadow showed over 40 degrees. This was because these contained ferromagnetic pigments such as an iron oxide, but those translational forces were very small owing to slight mass. However, it is desirable that these should be removed before MRI examination to prevent secondary problems.