不同股骨柄设计的体外界面及骨水泥壳分析

In vitro interface and cement mantle analysis of different femur stem designs.

作者信息

Gravius S, Wirtz D C, Siebert C H, Andereya St, Mueller-Rath R, Maus U, Mumme T

机构信息

Department of Orthopaedic Surgery, University Hospital Bonn, Sigmund-Freud-Strasse 25, 53125 Bonn, Germany.

出版信息

J Biomech. 2008;41(9):2021-8. doi: 10.1016/j.jbiomech.2008.03.041. Epub 2008 Jun 2.

Abstract

The aim of this study is to define stem design related factors causing both gaps in the metal-bone cement interface and cracks within the cement mantle. Six different stem designs (Exeter; Lubinus SP II; Ceraver Osteal; Mueller-straight stem; Centega; Spectron EF) (n=15 of each design) were cemented into artificial femur bones. Ten stems of each design were loaded, while five stems served as an unloaded control. Physiologically adapted cyclical loading (DIN ISO 7206-4) was performed with a hip simulator. After loading both interfaces and the bone cement itself were analysed regarding gaps and cracks in the cement mantle. Significant differences between the stem designs concerning gaps in the metal-bone cement interface and cracks in the cement mantle became apparent. Additionally, a high correlation between gaps in the metal-bone cement interface and cracks within the cement mantle could be proven. Gaps in the metal-bone cement interface but no cracks within the cement mantle were seen in the unloaded specimens. Differences between the unloaded control groups and the cyclical loaded stems regarding the longitudinal extension and width of gaps in the metal-bone cement interface were obvious. The designs of cemented femoral stems have an influence on both the quality of the metal-bone cement contact and the failure rate of the cement mantle. Less interface gaps and less cement defects were found with anatomically formed, collared, well-rounded stem designs without undercuttings.

摘要

本研究的目的是确定导致金属-骨水泥界面出现间隙以及骨水泥壳内出现裂缝的与柄部设计相关的因素。将六种不同的柄部设计(埃克塞特;卢比努斯SP II;塞拉弗骨型;穆勒直柄;森特加;光谱EF)(每种设计n = 15)用骨水泥固定到人工股骨中。每种设计的十个柄部进行加载,而五个柄部作为未加载的对照。使用髋关节模拟器进行生理适应性循环加载(DIN ISO 7206-4)。加载后,分析两个界面以及骨水泥本身在骨水泥壳中的间隙和裂缝情况。柄部设计在金属-骨水泥界面的间隙和骨水泥壳中的裂缝方面存在显著差异。此外,可以证明金属-骨水泥界面的间隙与骨水泥壳内的裂缝之间存在高度相关性。在未加载的标本中观察到金属-骨水泥界面有间隙,但骨水泥壳内无裂缝。未加载的对照组与循环加载的柄部在金属-骨水泥界面间隙的纵向延伸和宽度方面存在明显差异。骨水泥固定型股骨柄的设计对金属-骨水泥接触质量和骨水泥壳的失败率均有影响。在无切槽的解剖形状、有领圈、圆钝的柄部设计中,界面间隙和骨水泥缺陷较少。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验