Masi Laura, Casamassima Franco, Polli Caterina, Menichelli Claudia, Bonucci Ivano, Cavedon Carlo
Medical Physics Department, Casa di Cura S Chiara, Firenze, Italy.
Int J Radiat Oncol Biol Phys. 2008 Jul 1;71(3):926-33. doi: 10.1016/j.ijrobp.2008.03.006.
An analysis is performed of the setup errors measured by a kV cone beam computed tomography (CBCT) for intracranial stereotactic radiotherapy (SRT) patients immobilized by a thermoplastic mask and a bite-block and positioned using stereotactic coordinates. We evaluated the overall positioning precision and accuracy of the immobilizing and localizing systems. The potential of image-guided radiotherapy to replace stereotactic methods is discussed.
Fifty-seven patients received brain SRT. After a frame-guided setup, before each fraction (131 fractions), a CBCT was acquired and the detected displacements corrected online. Translational and rotational errors were analyzed calculating overall mean and standard deviation. A separate analysis was performed for bite-block (in conjunction with mask) and for simple thermoplastic mask. Interobserver variability for CBCT three-dimensional registration was assessed. The residual error after correction and intrafractional motion were calculated.
The mean module of the three-dimensional displacement vector was 3.0 +/- 1.4 mm. Setup errors for bite block and mask were smaller (2.9 +/- 1.3 mm) than those for thermoplastic mask alone (3.2 +/- 1.5 mm), but statistical significance was not reached (p = 0.15). Interobserver variability was negligible. The maximum margin calculated for residual errors and intra fraction motion was small but not negligible (1.57 mm).
Considering the detected setup errors, daily image guidance is essential for the efficacy of SRT treatments when mask immobilization is used, and even when a bite-block is used in conjunction. The frame setup is still used as a starting point for the opportunity of rotational corrections. Residual margins after on-line corrections must be evaluated.
对采用热塑性面罩和咬块固定并使用立体定向坐标定位的颅内立体定向放射治疗(SRT)患者,通过千伏锥形束计算机断层扫描(CBCT)测量的摆位误差进行分析。我们评估了固定和定位系统的整体定位精度和准确性。讨论了图像引导放射治疗取代立体定向方法的潜力。
57例患者接受脑部SRT。在框架引导摆位后,每次分次照射前(共131次分次照射),采集CBCT并在线校正检测到的位移。分析平移和旋转误差,计算总体均值和标准差。分别对咬块(与面罩联合使用)和单纯热塑性面罩进行分析。评估CBCT三维配准的观察者间变异性。计算校正后的残余误差和分次照射期间的运动。
三维位移向量的平均模为3.0±1.4mm。咬块和面罩的摆位误差(2.9±1.3mm)比单纯热塑性面罩的误差(3.2±1.5mm)小,但未达到统计学显著性(p = 0.15)。观察者间变异性可忽略不计。为残余误差和分次照射期间运动计算的最大边界较小但不可忽略(1.57mm)。
考虑到检测到的摆位误差,当使用面罩固定时,即使联合使用咬块,每日图像引导对于SRT治疗的疗效至关重要。框架摆位仍用作进行旋转校正的起点。必须评估在线校正后的残余边界。