Suppr超能文献

立体定向放射外科中 HyperArc 治疗单发和多发脑转移瘤的旋转摆位误差的剂量学效应。

Dosimetric effect of rotational setup errors in stereotactic radiosurgery with HyperArc for single and multiple brain metastases.

机构信息

Department of Radiation Oncology, Osaka International Cancer Institute, Osaka, Japan.

Department of Medical Physics and Engineering, Graduate School of Medicine, Osaka University, Suita, Japan.

出版信息

J Appl Clin Med Phys. 2019 Oct;20(10):84-91. doi: 10.1002/acm2.12716. Epub 2019 Sep 11.

Abstract

PURPOSE

In stereotactic radiosurgery (SRS) with single-isocentric treatments for brain metastases, rotational setup errors may cause considerable dosimetric effects. We assessed the dosimetric effects on HyperArc plans for single and multiple metastases.

METHODS

For 29 patients (1-8 brain metastases), HyperArc plans with a prescription dose of 20-24 Gy for a dose that covers 95% (D ) of the planning target volume (PTV) were retrospectively generated (Ref-plan). Subsequently, the computed tomography (CT) used for the Ref-plan and cone-beam CT acquired during treatments (Rot-CT) were registered. The HyperArc plans involving rotational setup errors (Rot-plan) were generated by re-calculating doses based on the Rot-CT. The dosimetric parameters between the two plans were compared.

RESULTS

The dosimetric parameters [D , D , D , homogeneity index, and conformity index (CI)] for the single-metastasis cases were comparable (P > 0.05), whereas the D for each PTV of the Rot-plan decreased 10.8% on average, and the CI of the Rot-plan was also significantly lower than that of the Ref-plan (Ref-plan vs Rot-plan, 0.93 ± 0.02 vs 0.75 ± 0.14, P < 0.01) for the multiple-metastases cases. In addition, for the multiple-metastases cases, the Rot-plan resulted in significantly higher V (P = 0.01), V (P = 0.02), V (P = 0.02), and V (P < 0.01) than those in the Ref-plan.

CONCLUSION

The rotational setup errors for multiple brain metastases cases caused non-negligible underdosage for PTV and significant increases of V to V in SRS with HyperArc.

摘要

目的

在立体定向放射外科(SRS)中,对于脑转移瘤的单次等中心治疗,旋转设置误差可能会导致相当大的剂量学影响。我们评估了旋转设置误差对单发性和多发性脑转移瘤 HyperArc 计划的剂量学影响。

方法

回顾性地为 29 名患者(1-8 个脑转移瘤)生成了 HyperArc 计划,处方剂量为 20-24 Gy,以覆盖计划靶区(PTV)的 95%(D )。随后,将用于 Ref-plan 的 CT 与治疗过程中采集的锥形束 CT(Rot-CT)进行配准。根据 Rot-CT 重新计算剂量,生成涉及旋转设置误差的 HyperArc 计划(Rot-plan)。比较了这两种计划的剂量学参数。

结果

在单发性转移瘤病例中,剂量学参数[D 、D 、D 、均匀性指数和适形性指数(CI)]具有可比性(P>0.05),但 Rot-plan 中每个 PTV 的 D 平均降低了 10.8%,且 Rot-plan 的 CI 也显著低于 Ref-plan(Ref-plan 与 Rot-plan,0.93±0.02 与 0.75±0.14,P<0.01)。此外,对于多发性转移瘤病例,Rot-plan 导致 PTV 的 V (P=0.01)、V (P=0.02)、V (P=0.02)和 V (P<0.01)显著升高。

结论

对于多发性脑转移瘤病例,旋转设置误差导致 SRS 中 PTV 剂量不足,并且 V 至 V 的显著增加。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验