Mödinger Yvonne, Anttila Eric D, Baker Grant M, Gross David C, Porporati Alessandro A
Medical Products Division, CeramTec GmbH, Plochingen, Germany.
MED Institute Inc., West Lafayette, IN, USA.
Arthroplast Today. 2023 Jul 23;22:101170. doi: 10.1016/j.artd.2023.101170. eCollection 2023 Aug.
Image artifacts caused by metal knee implants in 1.5T and 3T magnetic resonance imaging (MRI) systems complicate imaging-based diagnosis of the peri-implant region after total knee arthroplasty. Alternatively, metal-free knee prostheses could effectively minimize MRI safety hazards and offer the potential for higher quality diagnostic images.
A novel knee arthroplasty device composed of BIOLOX , an alumina matrix composite (AMC) ceramic, was tested in a magnetic resonance (MR) environment. American Society for Testing and Materials test methods were used for evaluating magnetically induced displacement force, magnetically induced torque, radiofrequency-induced heating, and MR image artifact.
Magnetically induced displacement force and magnetically induced torque results of the AMC ceramic knee indicated that these effects do not pose a known risk in a clinical MR environment, as assessed in a 3T magnetic field. Moreover, minimal radiofrequency-induced heating of the device was observed. In addition, the AMC ceramic knee demonstrated minimal MR image artifacts (7 mm) in comparison to a cobalt-chromium knee (88 mm). The extremely low magnetic susceptibility of AMC (2 ppm) underlines that it is a nonmetallic and nonmagnetic material well suited for the manufacturing of MR Safe orthopaedic implants.
The AMC ceramic knee is a novel metal-free total knee arthroplasty device that can be regarded as MR Safe, as suggested by the absence of hazards from the exposure of this implant to a MR environment. The AMC ceramic knee presents the advantage of being scanned with superior imaging results in 3T MRI systems compared to alternative metal implants on the market.
在1.5T和3T磁共振成像(MRI)系统中,金属膝关节植入物所导致的图像伪影使全膝关节置换术后植入物周围区域的基于成像的诊断变得复杂。另外,无金属膝关节假体可有效降低MRI安全风险,并有可能提供更高质量的诊断图像。
一种由BIOLOX(一种氧化铝基复合材料(AMC)陶瓷)组成的新型膝关节置换装置在磁共振(MR)环境中进行了测试。采用美国材料与试验协会的测试方法来评估磁致位移力、磁致扭矩、射频感应加热和MR图像伪影。
AMC陶瓷膝关节的磁致位移力和磁致扭矩结果表明,在3T磁场中评估时,这些效应在临床MR环境中不会构成已知风险。此外,观察到该装置的射频感应加热极小。此外,与钴铬合金膝关节(88mm)相比,AMC陶瓷膝关节的MR图像伪影极小(7mm)。AMC极低的磁化率(2ppm)表明它是一种非常适合制造MR安全骨科植入物的非金属和非磁性材料。
AMC陶瓷膝关节是一种新型的无金属全膝关节置换装置,由于该植入物暴露于MR环境中没有危害,因此可被视为MR安全的。与市场上的其他金属植入物相比,AMC陶瓷膝关节在3T MRI系统中进行扫描时具有成像结果更佳的优势。