Effect of undersizing on the long-term stability of the Exeter hip stem: A comparative in vitro study.

作者信息

Cristofolini Luca, Erani Paolo, Bialoblocka-Juszczyk Ewa, Ohashi Hirotsugu, Iida Satoshi, Minato Izumi, Viceconti Marco

机构信息

Laboratorio di Tecnologia Medica, Istituto Ortopedico Rizzoli, Via di Barbiano 1/10, 40136 Bologna, Italy.

出版信息

Clin Biomech (Bristol). 2010 Nov;25(9):899-908. doi: 10.1016/j.clinbiomech.2010.07.003. Epub 2010 Jul 24.

Abstract

BACKGROUND

Even for clinically successful hip stems such as the Exeter-V40 occasional failures are reported. It has been reported that sub-optimal pre-operative planning, leading to implant undersizing and/or thin cement mantle, can explain such failures. The scope of this study was to investigate whether stem undersizing and a thin cement mantle are sufficient to cause implant loosening.

METHODS

A comparative in vitro study was designed to compare hip implants prepared with optimal and smaller than optimal stem size. Exeter-V40, a highly polished cemented hip stem, was used in both cases. Tests were carried out simulating 24 years of activity of active hip patients. A multifaceted approach was taken: inducible and permanent micromotions were recorded throughout the test; cement micro-cracks were quantified using dye penetrants and statistically analyzed.

FINDINGS

The implants with an optimal stem size withstood the entire mechanical test, with low and stable inducible micromotions and permanent migrations during the test, and with moderate fatigue damage in the cement mantle after test completion. Conversely, the undersized specimens showed large and increasing micromotions, and failed after few loading cycles, because of macroscopic cracks in the proximal part of the cement mantle. While results for the optimal stem size are typical for stable hip stems, those for the undersize stem indicate a critical scenario.

INTERPRETATION

These results confirm that even a clinically successful hip prosthesis such as the Exeter-V40 is prone to early loosening if a stem smaller than the optimal size is implanted.