用于监测国际临床化学和检验医学联合会(IFCC)糖化血红蛋白参考测量程序与美国、日本和瑞典指定比较方法之间关系的统计方法。

Statistical methods for monitoring the relationship between the IFCC reference measurement procedure for hemoglobin A1c and the designated comparison methods in the United States, Japan, and Sweden.

作者信息

Geistanger Andrea, Arends Sabine, Berding Christoph, Hoshino Tadao, Jeppsson Jan-Olof, Little Randie, Siebelder Carla, Weykamp Cas

机构信息

Roche Diagnostics GmbH, Department of Biostatistics, Penzberg, Germany.

出版信息

Clin Chem. 2008 Aug;54(8):1379-85. doi: 10.1373/clinchem.2008.103556. Epub 2008 Jun 6.

Abstract

BACKGROUND

The American Diabetes Association (ADA)/European Association for the Study of Diabetes (EASD)/International Diabetes Federation (IDF)/IFCC Consensus Statement on the worldwide standardization of HbA(1c) states that "... [HbA(1c)] results are to be reported world-wide in IFCC units ... and derived NGSP units ... , using the IFCC-NGSP master equation."

METHODS

We describe statistical methods to evaluate and monitor the relationships as expressed in master equations (MEs) between the IFCC Reference Measurement procedure (IFCC-RM) and designated comparison methods (DCMs) [US National Glycohemoglobin Standardization Program (NGSP), Japanese Diabetes Society/Japanese Society for Clinical Chemistry (JDS/JSCC), and Mono-S in Sweden]. We applied these statistics, including uncertainty calculations, to 12 studies in which networks of reference laboratories participated, operating the IFCC-RM and DCMs.

RESULTS

For NGSP and Mono-S, slope, intercept, and derived percentage HbA(1c) at the therapeutic target show compliance with the respective MEs in all 12 studies. For JDS/JSCC, a slight deviation is seen in slope and derived percentage HbA(1c) in 2 of the 12 studies. Using the MEs, the uncertainty in an assigned value increases from 0.42 mmol/mol HbA(1c) (IFCC-RM) to 0.47 (NGSP), 0.49 (JDS/JSCC), and 0.51 (Mono-S).

CONCLUSIONS

We describe sound statistical methods for the investigation of relations between networks of reference laboratories. Application of these statistical methods to the relationship between the IFCC-RM and DCMs in the US, Japan, and Sweden shows that they are suitable for the purpose, and the results support the applicability of the ADA/EASD/IDF/IFCC Consensus Statement on HbA1c measurement.

摘要

背景

美国糖尿病协会(ADA)/欧洲糖尿病研究协会(EASD)/国际糖尿病联盟(IDF)/国际临床化学和检验医学联合会(IFCC)关于糖化血红蛋白(HbA₁c)全球标准化的共识声明指出:“……[HbA₁c]结果应在全球范围内以IFCC单位……以及导出的美国国家糖化血红蛋白标准化计划(NGSP)单位……报告,使用IFCC - NGSP主方程。”

方法

我们描述了统计方法,以评估和监测IFCC参考测量程序(IFCC - RM)与指定比较方法(DCMs)[美国国家糖化血红蛋白标准化计划(NGSP)、日本糖尿病学会/日本临床化学学会(JDS/JSCC)以及瑞典的Mono - S]之间主方程(MEs)所表达的关系。我们将这些统计方法,包括不确定度计算,应用于12项参考实验室网络参与的研究,这些研究操作IFCC - RM和DCMs。

结果

对于NGSP和Mono - S,斜率、截距以及治疗靶点处导出的HbA₁c百分比在所有12项研究中均符合各自的主方程。对于JDS/JSCC,在12项研究中的2项研究中观察到斜率和导出的HbA₁c百分比存在轻微偏差。使用主方程,赋值的不确定度从0.42 mmol/mol HbA₁c(IFCC - RM)增加到0.47(NGSP)、0.49(JDS/JSCC)和0.51(Mono - S)。

结论

我们描述了用于研究参考实验室网络之间关系的可靠统计方法。将这些统计方法应用于美国、日本和瑞典的IFCC - RM与DCMs之间的关系表明,它们适用于该目的,并且结果支持ADA/EASD/IDF/IFCC关于HbA₁c测量的共识声明的适用性。